07 | Keysight N1911A/N1912A P-Series Power Meters and N1921A/N1922A Wideband Power Sensors - Data Sheet  
1 mW Power Reference (continued)  
Physical characteristics  
Dimensions  
The following dimensions exclude front and rear panel protrusions:  
88.5 mm H x 212.6 mm W x 348.3 mm D (3.5 in x 8.5 in x 13.7 in)  
Command language  
Shipping weight  
Display  
N1911A ≤ 3.5 kg (7.7 lb) approximate  
N1912A ≤ 3.7 kg (8.1 lb) approximate  
N1911A ≤ 7.9 kg (17.4 lb) approximate  
N1912A ≤ 8.0 kg (17.6 lb) approximate  
3.8 inch TFT Color LCD  
Environmental conditions  
General  
The following dimensions exclude front and rear panel protrusions:  
88.5 mm H x 212.6 mm W x 348.3 mm D (3.5 in x 8.5 in x 13.7 in)  
Operating  
Temperature  
0 °C to 55 °C  
Maximum humidity  
Maximum altitude  
95 % at 40 °C (non-condensing)  
3,000 meters (9,840 feet)  
Storage  
Non-operating storage temperature  
Non-operating maximum humidity  
Non-operating maximum altitude  
–40 °C to +70 °C  
90 % at 65 °C (non-condensing)  
15,420 meters (50,000 feet)  
System Specifications and Characteristics  
The video bandwidth in the meter can be set to High, Medium, Low and Off. The video bandwidths stated in the table below are not the  
3 dB bandwidths, as the video bandwidths are corrected for optimal flatness (except the Off filter). Refer to Figure 2 for information on  
the flatness response. The Off video bandwidth setting provides the warranted rise time and fall time specification and is the recom-  
mended setting for minimizing overshoot on pulse signals.  
Dynamic response - rise time, fall time, and overshoot versus video bandwidth settings  
Video bandwidth setting  
Parameter  
Off  
Low: 5 MHz  
Medium: 15 MHz  
High: 30 MHz  
< 500 MHz  
< 36 ns  
< 5 %  
> 500 MHz  
≤ 13 ns  
< 5 %  
Rise time/fall time 1  
Overshoot 2  
< 56 ns  
< 25 ns  
≤ 13 ns  
For Option 107 (10 m cable), add 5 ns to the rise time and fall time specifications.  
1. Specified as 10% to 90% for rise time and 90% to 10% for fall time on a 0 dBm pulse.  
2. Specified as the overshoot r For Option 107 (10 m cable), add 5 ns to the rise time and fall time specifications. elative to the settled pulse top power.  
Recorder output and video output  
The recorder output is used to output the corresponding voltage for the measurement a user sets on the Upper/Lower  
window of the power meter.  
The video output is the direct signal output detected by the sensor diode, with no correction applied. The video output provides a DC  
voltage proportional to the measured input power through a BNC connector on the rear panel. The DC voltage can be displayed on an  
oscilloscope for time measurement. This option replaces the recorder output on the rear panel. The video output impedance is 50 ohm.