04 | Keysight N1911A/N1912A P-Series Power Meters and N1921A/N1922A Wideband Power Sensors - Data Sheet
P-Series Power Meter Specifications
Meter uncertainty
Instrumentation linearity
± 0.8 %
Timebase
Timebase range
Accuracy
Jitter
2 ns to 100 msec/div
±10 ppm
≤ 1 ns
Zero Set
Zero Set (CW)
Zero Set (Peak)
0.0000175%
0.015%
Trigger
Internal trigger
Range
–20 to +20 dBm
0.1 dB
Resolution
Level accuracy
± 0.5 dB
1
Latency
160 ns ± 10
≤ 5 ns rms
Jitter
External TTL trigger input
High
Low
> 2.4 V
< 0.7 V
2
Latency
30 ns ± 10 ns
50 Ω
Impedance
Jitter
≤ 5 ns rms
Trigger delay
Delay range
± 1.0 s, maximum
Delay resolution
Trigger hold-off
Range
1 % of delay setting, 10 ns maximum
1 μs to 400 ms
Resolution
1 % of selected value (to a minimum of 10 ns)
Trigger level threshold hysteresis
Range
± 3 dB
Resolution
0.05 dB
1. Internal trigger latency is defined as the delay between the applied RF crossing the trigger level and the meter switching into the triggered state.
2. External trigger latency is defined as the delay between the applied trigger crossing the trigger level and the meter switching into the triggered state.
3. External trigger output latency is defined as the delay between the meter entering the triggered state and the output signal switching.