08 | Keysight | IntegraVision PA2200 Series Power Analyzers - Data Sheet
See it, measure it, prove it.
Main screen
Transient measurements
See real-time voltage, current and power on all channels
simultaneously, with each parameter on its own grid (the default
screen layout). Screen layout and functions are the same as a
traditional oscilloscope.
Use single-shot measurement mode to analyze power transients.
Here the Keysight IntegraVision power analyzer displays an AC
line voltage full-cycle dropout on one channel (yellow trace) and
the corresponding response of a DC power supply output on the
other channel (green trace). Thanks to single-shot measurement
and visualization, you can easily see that the 20 V DC output
drops about 0.6 V and then recovers in 10 ms.
Power analysis panel
Measurement with cursors
Use cursors to make common measurements, shown in the right
side panel. Here, the Keysight IntegraVision power analyzer
displays in-rush current (middle current trace) and power (bottom
power trace) following an AC line input voltage turn-on (top
voltage trace). The cursors show a peak of 6.75 A during the first
cycle.
Press the [Analyze] key to view power quality measurements,
shown in the blue Power Quality analysis panel. While displaying
live power quality measurements, you can also view real-time
voltage, current and power waveforms.