06 | Keysight | FieldFox Handheld Analyzers 4/6.5/9/14/18/26.5/32/44/50 GHz - Data Sheet
Cable and Antenna Analyzer and Vector Network Analyzer (continued)
Test port output specifications (continued)
Test port output power (dBm), high power
N995xA
Typical
Nominal
Port 1
Port 2
300 kHz to 2 MHz
> 2 MHz to 1 GHz
> 1 to 6.5 GHz
0
2
0
2
—
—
2
0
—
> 6.5 to 18 GHz
4
1
—
> 18 to 39 GHz
1
-2
-5
-7
—
> 39 to 46 GHz
-2
-4
—
—
> 46 to 50 GHz
Test port output power (dBm), low power
N995xA
Typical
Nominal
Port 1
-35
Port 2
-38
500 kHz to 10 MHz
> 10 MHz to 10 GHz
> 10 to 20 GHz
—
-38
-42
—
-43
-47
—
> 20 to 44 GHz
-44
-50
—
—
> 44 to 50 GHz
-53
-55
Max leveled output power (dBm)
N995xA
Typical
Nominal
Port 1
Port 2
-2
500 kHz to 10 MHz
> 10 MHz to 25 GHz
> 25 to 32 GHz
-2
0
—
—
—
—
—
0
0
-4
> 32 to 44 GHz
-3
-7
-6
> 44 to 50 GHz
-10
Output power range
CAT
High, low, and manual. Default (preset) power is high
Manual power is flattened.
VNA
High, low, and manual. Default (preset) power is manual, −15 dBm.
Manual power is flattened.
Power step size
Power settable in 1 dB steps across power range. Flat power, in 1 dB steps, is available across the
whole frequency span, nominal
Power level accuracy1
N991xA, N992xA
N995xA
Typical
± 1.5 dB at −15 dBm, for frequencies > 250 kHz
± 0.7 dB at -15 dBm, for frequencies > 500 kHz to 10 MHz
± 0.5 dB at -15 dBm, for frequencies > 10 MHz to 50 GHz
Power level linearity
N995xA
Nominal
Port 1 or port 2, −25 dBm ≤ P < max leveled power each port
10 MHz to 50 GHz
± 0.5 dB
1. N991xA and N992xA power levels are calibrated in the factory using a broadband power sensor, which means all tones (fundamental and harmonics) are
included. N995xA power levels are calibrated based on PNA-X’s tuned receiver, which means primarily the fundamental is included (for frequency ≥ 10 MHz).