12 | Keysight | FieldFox Handheld Analyzers 4/6.5/9/14/18/26.5/32/44/50 GHz - Data Sheet
Corrected Measurement Uncertainty for N9913A/14A/15A/16A/17A/18A and
N9925A/26A/27A/28A (continued)
Full 2-port calibration, 85054D Type-N (m) calibration kit, spec
Power level of -15 dBm, 10 Hz IF bandwidth, no averaging, battery saver off, and 60-minute warm-up time. Includes uncertainties due
to drift, noise, compression, and dynamic accuracy. Coverage factor of x1 applied to uncertainties, for ease of comparison with other
industry handheld analyzers.
Corrected performance (dB)
Directivity
≤ 0.5 GHz
40
0.5 to 2 GHz
40
2 to 8 GHz
36
8 to 18 GHz
34
Source match
38
33
33
27
Load match
37
35
32
27
Reflection tracking
Transmission tracking
± 0.006
± 0.070
± 0.006
± 0.100
± 0.009
± 0.150
± 0.027
± 0.430
Transmission uncertainty (S21, S12)
Reflection uncertainty (S11, S22)