37 | Keysight | FieldFox Handheld Analyzers 4/6.5/9/14/18/26.5/32/44/50 GHz - Data Sheet  
Extended Range Transmission Analysis (ERTA) (continued)  
Dynamic range and maximum attenuation  
Dynamic range is the difference between the maximum output power available from FieldFox’s source and the noise floor of the sec-  
ond FieldFox, while ensuring that neither FieldFox’s ADC goes into over-range. Dynamic range also accounts for the loss of the power  
splitter. Dynamic range is applicable when testing devices such as filters, where there is low loss in the passband, and significant loss  
in the stopband, and both passband and stopband need to be on the display at the same time (same sweep).  
Maximum attenuation is the difference between maximum output power available from FieldFox’s source and the noise floor of  
FieldFox. It also accounts for the loss of power splitter. Maximum attenuation is applicable when testing devices such as cables, which  
have relatively uniform loss over the swept frequency range.  
The values shown are based on the recommended minimum RBW of 3 kHz when the frequency references are locked via GPS, and 300  
kHz when the frequency references are unlocked. Locking the frequency references to GPS allows for greater frequency accuracy of  
the FieldFoxes and use of a narrower RBW, which in turn results in a lower DANL, and hence a wider measurement range. When the  
GPS signals cannot be present at all times, the GPS hold-over mode can be used.  
Dynamic range (dB)  
N991xA, N993xA  
Typical  
Preamp off  
Preamp on  
Preamp off  
Preamp on  
Frequency references  
locked to GPS, RBW 3 kHz  
Frequency references  
locked to GPS, RBW 3 kHz  
Frequency references  
unlocked, RBW 300 kHz  
Frequency references  
unlocked, RBW 300 kHz  
> 2 MHz1 to 6 GHz  
> 6 to 13 GHz  
> 13 to 22 GHz  
> 22 to 25 GHz  
> 25 to 26.5 GHz  
88  
86  
70  
63  
58  
83  
83  
86  
83  
77  
68  
66  
50  
43  
38  
63  
63  
66  
63  
57  
Maximum attenuation (dB)  
Typical  
N991xA, N993xA  
Preamp off  
Preamp on  
Preamp off  
Preamp on  
Frequency references  
locked to GPS, RBW 3 kHz  
Frequency references  
locked to GPS, RBW 3 kHz  
Frequency references  
unlocked, RBW 300 kHz  
Frequency references  
unlocked, RBW 300 kHz  
> 2 MHz to 6 GHz  
> 6 to 13 GHz  
> 13 to 22 GHz  
> 22 to 25 GHz  
> 25 to 26.5 GHz  
93  
86  
70  
63  
58  
108  
103  
91  
83  
77  
73  
66  
50  
43  
38  
88  
83  
71  
63  
57  
1. Dynamic range is decreased from 3 to 9 dB at 2 MHz.  