Spectrum Analyzer (continued)
Description
Specification
Supplemental Information
Sweep Acquisition, Span > 0 Hz
Range
1 to 5000. Number of data acquisitions per
trace point. Value is normalized to the
minimum required to achieve amplitude
accuracy with CW signals.
Auto coupled. For pulsed RF signals,
manually increase the sweep acquisition
value to maximize the pulse spectrum
envelope.
Resolution
Readout
1
Measured value representing time required to tune receiver, acquire data, and process
trace.
Trigger
Type
Free Run, Video, External
Positive, Negative edge
Range: 0 to 10 sec
Slope
Delay
Range
Resolution: 100 nsec
Auto Trigger
Forces a periodic acquisition in the absence of a trigger event
Range: 0 sec (OFF) to 10 sec
Time Gating
Gate Method
Triggered FFT
Gate Delay Range
Same as Trigger Delay
Trace Update
Span = 20 MHz, RBW = 3 kHz 1
Span = 100 MHz, RBW auto coupled 2
Span = 6 GHz, RBW auto coupled 3
5.9 updates/s (nominal)
16.7 updates/s (nominal)
1.7 update/s (nominal)
Trace Points
101, 201, 401, 601, 801, 1001 (Defaults to 401)
1.
2.
3.
1.5 updates/s; applicable for N9912A with serial number prefix <MY5607/SG5607/US5607 and N9912A not upgraded with Option
N9910HU-500.
7 updates/s; applicable for N9912A with serial number prefix <MY5607/SG5607/US5607 and N9912A not upgraded with Option
N9910HU-500.
1 update/s; applicable for N9912A with serial number prefix <MY5607/SG5607/US5607 and N9912A not upgraded with Option
N9910HU-500.
16