04 | Keysight | N8834A MultiScope Application - Data Sheet
MultiScope Software Architecture
The leader and the followers
Control PC
The leader oscilloscope connects to the control PC via LAN or
USB or can act as the controller in the absence of the PC. The
one leader oscilloscope can connect two to nine followers. The
connection between oscilloscopes is daisy chained with cables
and splitters.
– Support edge or glitch trigger on leader (other triggering
modes (including InfiniiScan) are not supported).
– One-time deskew eliminates frame-to-frame skew created by
trigger daisy chain connection.
The N8834A MultiScope application runs on a Windows 7 PC.
All the scopes are connected to the control PC via LAN or USB.
A control PC must be used if InfiniiVision scopes exist in the
multiscope configuration.
– PC supports both hosted/online and offline analysis.
– Scope front panels are disabled during MultiScope operation.
MultiScope is controlled by a PC installed with the N8900A
Infiniium Offline analysis software or the leader scope in the
absence of a PC.
– Achieve inter-scope intrinsic jitter of 150 fs with precision
calibration on the 90000 Q-Series and Z-Series Infiniium
oscilloscopes.
– Display up to 16 grids per window, up to 8 windows.
– Full support of all measurements, functions, and analysis on
all channels.
– 10-MHz reference clock aligns sample for deep memory
synchronization.
– Test at the comfort of your workbench.
Daisy chained connection with power splitter and BNC cables.
View of 6 channels and 3 eye diagrams on a single window with 8 grids.
Leader scope is used to control and display multichannel waveforms.