25 | Keysight | Infiniium 9000 Series Oscilloscopes - Data Sheet
Infiniium 9000 Series Performance Characteristics (Continued)
Display
Display
15 inch color XGA TFT-LCD with touch screen
Display intensity grayscale
– Resolution
– Annotation
– Grids
64-level intensity-graded display
1024 pixels horizontally x 768 pixels vertically
Up to 12 labels, with up to 100 characters each, can be inserted into the waveform area
Can display 1, 2 or 4 waveform grids
– Waveform styles
Connected dots, dots, variable persistence, infinite persistence, color graded infinite persistence. Includes up to 64
levels of intensity-graded waveform
Waveform update rate (10 GS/s, Segmented mode: Maximum up to 250,000 waveforms/sec
50 ns/div, sin(x)/x: on)
Real time mode:
– Typical of 700 waveforms/sec with 1 kpts memory
– Typical of 230 waveforms/sec with 100 kpts memory
– Typical of 130 waveforms/sec with 1 Mpts memory
Computer system and peripherals, I/O ports
Computer system and peripherals
– Operating system
– CPU
Windows 7 Embedded Standard
Intel® Core 2 Duo, M890, 3.0 GHz microprocessor
– PC system memory
– Drives
4 GB
≥ 250-GB internal hard drive (optional removable hard drive)
– Peripherals
Optical USB mouse and compact keyboard supplied. All Infiniium models support any Windowscompatible input device
with a PS/2 or USB interface.
File types
– Waveforms
Compressed internal format (*.wfm), comma separated values (*.csv), .hdf5, .bin, tab separated values (*.tsv), ability to
save .osc (composite including both setup and waveform. and Y value files (*.txt)
BMP, TIFF, GIF, PNG or JPEG
– Images
I/O ports
– LAN
RJ-45 connector, supports 10Base-T, 100Base-T, and 1000Base-T. Enables Web-enabled remote control, e-mail on
trigger, data/file transfers and network printing
– RS-232 (serial)
– PS/2
9-pin, COM1, printer and pointing device support
Two ports. Supports PS/2 pointing and input devices.
– USB 2.0 Hi-Speed
Three 2.0 high-speed ports on front panel plus four ports on side panel. Allows connection of USB peripherals like
storage devices and pointing devices while the oscilloscope is on. One device port on side for instrument control.
15 pin XGA on side of scope, full output of scope display or dual monitor video output, DVI
DC (± 2.4 V); square wave ~755 Hz with ~200 ps rise time
– Video output
– Auxiliary output
Time base reference output
10 MHz, Amplitude into 50 ohms: 800 mV pp to 1.26 V pp (4 dBm ± 2 dB) if derived from internal reference. Tracks
external reference input amplitude ± 1 dB if applied and selected.
Time base reference output
LXI compliance
Must be 10 MHz, input Z = 50 ohms. Minimum 500 mV pp (–2 dBm), maximum 2.0 V pp (+10 dBm)
LXI Class C