25 | Keysight | Infiniium V-Series Oscilloscopes - Data Sheet  
Achieve Clarify Faster - Frequency Domain Performance (Continued)  
Trying to interpret traditional oscilloscope time-domain specifications can be challenging when trying to determine if a specific scope  
can be recommended for RF, µW, and mmW measurements� With correction filters, low-noise front end and spurious free dynamic  
range calibration, the V-Series oscilloscopes can be used for wideband frequency domain applications�  
V-Series typical values (tested at 33 GHz bandwidth, 1  
channel on one scope unless noted)  
Sensitivity/noise density (1 mV/div; —38 dBm range)  
Power spectral density measurement at 1.0001 GHz, 1.0001 GHz center frequency,  
500 kHz span, and 3 kHz RBW  
159 dBm/Hz  
Noise figure (derived from measurement above)  
+15 dB  
Signal-to-noise ratio/dynamic range  
(—1 dBm, 1 GHz input carrier, 0 dBm scope input range)  
1 GHz center frequency, 100 MHz span, 1 kHz RBW, measurement at +20 MHz from  
center  
+111 dB  
Absolute amplitude accuracy (0 to 30 GHz)  
Deviation from linear phase (0 to 33 GHz)  
± 0.5 dB  
± 3 deg  
Phase noise (at 1 GHz)  
10 kHz offset  
100 kHz offset  
—125 dBc/Hz  
—131 dBc/Hz  
EVM  
(802.121 2.4 GHz carrier, 20 MHz wide, 64 QAM)  
—47 dB (0.47%)  
Spurious responses (—4.6 dBm input signal, —4 dBm input range)  
Spurious free dynamic range (SFDR)  
1 GHz, —4.6 dBm signal present at input, FFT = 5 GHz span, 3 GHz center, 100 kHz RBW  
+67 dB  
2nd harmonic distortion  
1 GHz input, —4.6 dBm, 5 GHz span, 3 GHz center, 100 KHz RBW  
—51 dBc  
—51 dBc  
3rd harmonic distortion  
1 GHz input, —4.6 dBm, 5 GHz span, 3 GHz center, 100 KHz RBW  
Two-tone third-order Intercept (TOI) Point  
0 dBm input tones, 2.436 and 2.438 GHz, 2 MHz separation, 2.437 GHz center frequency, +26 dBm  
10 MHz span, 30 kHz RBW, 8 dBm range  
Input match (S11)  
(< 50 mV/div, 0 to 30 GHz, no attenuation)  
(> = 50 mV/div, 0 to 30 GHz, no attentuation)  
—15 dB; 1.4 VSWR  
—21 dB; 1.2 VSWR  
Typical frequency domain characteristics for the Infiniium V-Series (not guaranteed, subject to change)�  
Using data acquired from the V-Series, VSA shows an excellent third  
order intercept (TOI) value of 26 dBm�  
Using data acquired from the V-Series, VSA shows an error vector  
magnitude (EVM) for IEEE 802�11 QAM 64 of 0�47%�  