06 | Keysight N1911A/N1912A P-Series Power Meters and N1921A/N1922A Wideband Power Sensors - Data Sheet  
1 mW Power Reference  
Note: The 1 mW power reference is provided for calibration of E-Series, 8480 Series and N8480 Series sensors.  
The P-Series sensors are automatically calibrated and therefore do not need this reference for calibration  
Power output  
1.00 mW (0.0 dBm). Factory set to ± 0.4 % traceable to the National Physical Laboratory  
Accuracy (over 2 years)  
±1.2 % (0 to 55 ºC)  
±0.4 % (25 ± 10 ºC)  
Frequency  
50 MHz nominal  
SWR  
1.08 (0 to 55 ºC)  
1.05 typical  
Connector type  
Type N (f), 50 Ω  
Rear-panel inputs/outputs  
Recorder output  
Analog 0-1 Volt, 1 kΩ output impedance, BNC connector. For dual-channel instruments there will be two  
recorder outputs  
GPIB, 10/100BaseT LAN and USB2.0  
Ground  
Interfaces allow communication with an external controller  
Binding post, accepts 4 mm plug or bare-wire connection  
Input has TTL compatible logic levels and uses a BNC connector  
Output provides TTL compatible logic levels and uses a BNC connector  
Trigger input  
Trigger output  
Line power  
Input voltage range  
Input frequency range  
Power requirement  
90 to 264 Vac, automatic selection  
47 to 63 Hz and 440 Hz  
N1911A not exceeding 50 VA (30 Watts)  
N1912A not exceeding 75 VA (50 Watts)  
Remote programming  
Interface  
GPIB interface operates to IEEE 488.2 and IEC65  
10/100BaseT LAN interface  
USB 2.0 interface  
Command language  
GPIB compatibility  
SCPI standard interface commands  
SH1, AH1, T6, TE0, L4, LE0, SR1, RL1, PP1, DC1, DT1, C0  
Measurement speed  
Measurement speed via remote  
interface  
≥ 1500 readings per second  
Regulatory information  
Electromagnetic compatibility  
Complies with the following requirements:  
IEC 61326-1:2005/EN 61326-1:2006  
CISPR11:2003/, EN 55011:1998+A1:1999+A2:2002 Group 1 Class A  
Canada: ICES/NMB-001:Issue 4, June 2006  
Australia/New Zealand: AS/NZS CISPR 11:2004  
Product safety  
Conforms to the following product specifications:  
IEC 61010-1:2010/EN 61010-1:2010 (3rd Edition)  
Canada: CAN/CSA-C22.2 No. 61010-1-12  
USA: ANSI/UL 61010-1:2012