11 | Keysight N1911A/N1912A P-Series Power Meters and N1921A/N1922A Wideband Power Sensors - Data Sheet  
Graphical Example  
A. System contribution to measurement uncertainty versus power level (equates to step 6 result/2)  
System uncertainty contribution - 1 sigma (%)  
100.0%  
N1921A: 500 MHz to 10 GH  
N1922A:18 to 40 GHz  
Other bands  
10.0%  
1.0%  
Note: This graph is valid for condi-  
tions of free-run operation, with a  
signal within the video bandwidth  
setting on the system.  
-35 -30 -25 -20 -15 -10 -5  
0
5
10 15 20  
Power (dBm)  
Humidity < 70%.  
B. Standard uncertainty of mismatch.  
Standard uncertainty of mismatch - 1 sigma (%)  
SWR  
1.0  
ρ
SWR  
1.8  
ρ
0.5  
0.45  
0.4  
0.00  
0.02  
0.05  
0.07  
0.09  
0.11  
0.13  
0.15  
0.17  
0.18  
0.20  
0.23  
0.26  
0.29  
0.31  
0.33  
0.35  
0.38  
0.39  
0.41  
0.43  
0.44  
0.46  
0.47  
0.49  
0.50  
1.05  
1.10  
1.15  
1.20  
1.25  
1.30  
1.35  
1.40  
1.45  
1.5  
1.90  
2.00  
2.10  
2.20  
2.30  
2.40  
2.50  
2.60  
2.70  
2.80  
2.90  
3.00  
0.35  
0.3  
0.25  
0.2  
0.15  
0.1  
0.05  
0
1.6  
1.7  
0
0.1  
0.2  
0.3  
0.4  
0.5  
r
DUT  
Note: The above graph shows the standard uncertainty of mismatch = ρDUT. ρSensor / ←2, rather than the mismatch uncertainty  
limits. This term assumes that both the source and load have uniform magnitude and uniform phase probability distributions.  
C. Combine A & B  
(Value from Graph A)2+ (Value from Graph B)2  
UC  
=
Expanded Uncertainty, k = 2, = 2. UC = . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
±
%