06 | Keysight | Infiniium DCA-X 86100D Wide-Bandwidth Oscilloscope Mainframe and Modules - Data Sheet  
Specifications (Continued)  
Mainframe specifications (Continued)  
86100D with option STR  
86100D with option ETR or with options ETR + PTB  
External divided trigger (AC coupled) Not available with option STR  
Input frequency range  
N/A  
50 MHz - 13 GHz (effective divide by 2)  
50 MHz - 15 GHz (characteristic)  
Option PTB: 50 MHz - 32 GHz (effective divide by 1)  
50 MHz - 13 GHz  
Pattern lock frequency range  
N/A  
50 MHz - 15 GHz (characteristic)  
Option PTB: 50 MHz - 32 GHz  
Pattern lock pattern length range  
Trigger sensitivity  
N/A  
N/A  
N/A  
1 to 2^23 UI in Scope and Eye/Mask Modes  
1 to 2^16 UI in Jitter Mode  
Slew rate > 2 V/ns:  
200 mVpp  
Max Trigger input: 2 Vpp  
Maximum input signal  
Option PTB: Max PTB Input: 1.3 Vpp with DC offset up  
to + 200 mV  
System jitter  
N/A  
Tested at 10 GHz, 200 mVpp  
1.2 ps rms for time delays less than 100 ns, (Maximum)  
750 fs rms at minimum delay setting, (characteristic)  
Option PTB when PTB is enabled:  
Condition 1: 2.4 GHz — < 4.0 GHz trigger, tested at  
2.4 GHz, 750 mVpp  
≤ 200 fs (characteristic)  
< 400 fs, with 54XXX, 8348X, or N1045A (non Option  
LOJ) module (characteristic)  
Condition 1: 4 GHz — 9.0 GHz trigger, tested at 5 GHz,  
750 mVpp  
≤ 120 fs (characteristic)  
< 400 fs, with 54XXX, 8348X, or N1045A (non Option  
LOJ) module (characteristic)  
Condition 1: > 9.0 GHz — 44.0 GHz trigger, tested at 10,  
20, and 40 GHz, 500 mVpp  
≤ 90 fs (characteristic)  
< 200 fs, with 54XXX, 8348X, or N1045A  
(non Option LOJ) module(characteristic)  
Inputs  
Trigger input connector  
2.92 mm (male)4  
2.92 mm (male)4  
Option PTB: PTB input: 2.92 mm (male)4  
50 Ω  
Trigger input nominal impedance  
50Ω  
Option PTB: PTB input: 50 Ω  
1. The internal precision timebase works with typical digital clock signals, such as a BERT output, as well as sine waves. If the rise time or fall time of the  
clock signal is less than 15% of the period of the clock (for example, less than 15 ps for a 10 GHz clock), reduce the edge speed by using an external  
low-pass filter. or length of cable. For the lowest jitter, use a signal that is as close as possible to the maximum signal amplitude (1.3 Vpp) and minimize any  
sub-harmonics.  
2. High Sensitivity Hysteresis mode improves the high frequency trigger sensitivity but is not recommended when using noisy, low frequency signals that may  
result in false triggers without normal hysteresis enabled.  
3. Module triggering enables precision waveform analyzer modules (86108), clock recovery modules (83496) and TDR modules to be used as a trigger source.  
Without support for module triggering, external connections between the module and the mainframe may be required.  
4. 86100D mainframes ship with a 2.92mm (f) to (f) adapter for this port.