欢迎访问ic37.com |
会员登录 免费注册
发布采购

N1076A-232 参数 Datasheet PDF下载

N1076A-232图片预览
型号: N1076A-232
PDF下载: 下载PDF文件 查看货源
内容描述: [Wide-Bandwidth Oscilloscope Mainframe and Modules]
分类和应用:
文件页数/大小: 35 页 / 3883 K
品牌: KEYSIGHT [ Keysight Technologies ]
 浏览型号N1076A-232的Datasheet PDF文件第1页浏览型号N1076A-232的Datasheet PDF文件第2页浏览型号N1076A-232的Datasheet PDF文件第3页浏览型号N1076A-232的Datasheet PDF文件第5页浏览型号N1076A-232的Datasheet PDF文件第6页浏览型号N1076A-232的Datasheet PDF文件第7页浏览型号N1076A-232的Datasheet PDF文件第8页浏览型号N1076A-232的Datasheet PDF文件第9页  
04 | Keysight | Infiniium DCA-X 86100D Wide-Bandwidth Oscilloscope Mainframe and Modules - Data Sheet  
Specifications  
General and mainframe  
Comparing specifications  
Specifications describe warranted performance over the  
temperature range of +10 to +40 °C (unless otherwise noted).  
The specifications are applicable for the temperature after the  
instrument is turned on for one hour, and while self-calibration  
is valid. Many performance parameters are enhanced through  
frequent, simple user calibrations. Characteristics provide useful,  
non-warranted information about the functions and performance  
of the instrument. Characteristics are printed in italic typeface.  
Product specifications and descriptions in this document are  
subject to change without notice.  
When comparing performance attributes between instruments,  
ensure you compare the same type of parameter. For example,  
compare warranted specifications from both instruments, or  
compare characteristics or typical performance. Warranted  
specifications include measurement uncertainties and are  
conservative compared to other types of unwarranted attributes.  
Factory calibration cycle - For optimum performance, the  
instrument should have a complete verification of specifications  
once every twelve months.  
General specifications  
Temperature  
Operating  
10 to +40 °C (50 to +104 °F)  
Non-operating  
Operating  
100/120 Vac 50/60/400 Hz  
–40 to +65 °C (–40 to +158 °F)  
Up to 4,600 meters (15,000 feet)  
With typical modules: 150 to 230 VA at 25 °C  
Altitude  
Power (max)  
220/240 Vac 50/60 Hz, 700 W maximum  
Mainframe without modules  
Typical module  
Weight  
20.5 kg (43 lbs)  
1.2 kg (2.6 lbs)  
Mainframe dimensions (excluding handle) Without front connectors and rear feet  
221 mm (H) x 426 mm (W) x 530 mm (D)  
(7 in x 16.76 in x 20.9 in)  
With front connectors and rear feet  
234 mm (H) x 426 mm (W) x 601 mm (D)  
(9.23 in x 16.76 in x 23.67 in)  
With front cover and rear feet  
234 mm (H) x 426 mm (W) x 612 mm (D)  
(9.23 in x 16.76 in x 24.1 in)  
Max relative humidity  
80% for temperatures up to 31 °C decreasing linearly to 50% relative humidity at 40 °C  
Supply voltage variation  
86100D system can operate with supply voltage fluctuations up to ± 10% of the nominal voltage  
Mainframe specifications  
Horizontal system (time base)  
Pattern lock  
Scale factor (full scale is ten divisions)  
Minimum  
Maximum  
Delays 1  
2 ps/div (with 86107A or 86100D-PTB: 500 fs/div)  
1 s/div  
250 ns/div  
Minimum  
24 ns  
40.1 ns default, 24 ns min  
Maximum  
1000 screen diameters or 10 s whichever is smaller  
1 ps + 1.0% of Δ time reading or 8 ps, whichever is smaller  
Time interval accuracy - jitter mode operation 1 ps  
≤ (screen diameter)/(record length) or 62.5 fs, whichever is  
larger  
1000 screen diameters or 25.401 μs whichever is smaller  
Time interval accuracy 2,3  
Jitter mode operation 4  
Time interval resolution  
Display units  
Bits or time (TDR mode–meters)  
Vertical system (channels)  
Number of channels  
Vertical resolution  
Full resolution channel  
Adjustments  
16 (simultaneous acquisition)  
54xxx, 83xxx, 86xxx: 14 bit A/D converter (16 or more bits with averaging). N10xx: 16 bit A/D converter  
Adjusts in a 1-2-5-10 sequence for coarse adjustment or fine adjustment resolution from the front panel knob  
Scale, offset, activate filter, sampler bandwidth, attenuation factor, transducer conversion factors  
Legacy UI: 16 to 16 K samples  
Record length  
FlexDCA without pattern lock: 16 to 64 K samples  
FlexDCA with pattern lock: 16 up to 32 M samples 5  
1. Time offset relative to the front panel trigger input on the instrument mainframe.  
2. Dual marker measurement performed at a temperature within 5 °C of horizontal calibration temperature.  
3. The maximum delay setting is 100 ns and delta time does not span across (28 + Nx4) ns ± 100 ps delay setting, where N=0,1,2….18. If delta time  
measurement span exceeds above criteria, time interval accuracy is 8 ps +0.1% of Δ time reading.  
4. Characteristic performance. Test configuration: PRBS of length 127 bits, Data and Clock 10 Gb/s.  
5. Maximum number of samples depends on pattern, number of active channels and available memory.  
 复制成功!