欢迎访问ic37.com |
会员登录 免费注册
发布采购

M9393A-FRX 参数 Datasheet PDF下载

M9393A-FRX图片预览
型号: M9393A-FRX
PDF下载: 下载PDF文件 查看货源
内容描述: [M9393A PXIe Performance Vector Signal Analyzer]
分类和应用:
文件页数/大小: 39 页 / 3429 K
品牌: KEYSIGHT [ Keysight Technologies ]
 浏览型号M9393A-FRX的Datasheet PDF文件第1页浏览型号M9393A-FRX的Datasheet PDF文件第2页浏览型号M9393A-FRX的Datasheet PDF文件第3页浏览型号M9393A-FRX的Datasheet PDF文件第5页浏览型号M9393A-FRX的Datasheet PDF文件第6页浏览型号M9393A-FRX的Datasheet PDF文件第7页浏览型号M9393A-FRX的Datasheet PDF文件第8页浏览型号M9393A-FRX的Datasheet PDF文件第9页  
04 | Keysight | M9393A PXIe Performance Vector Signal Analyzer – Data Sheet  
Recommended best practices in use  
Technical Specifications and  
Characteristics  
Use slot blockers and EMC filler panels in empty module  
slots to ensure proper operating temperatures. Keysight  
chassis and slot blockers optimize module temperature  
performance and reliability of test.  
Definitions for specifications  
Temperatures referred to in this document are defined as  
follows:  
– Set chassis fan to high at environmental temperatures  
above 45 °C.  
Full temperature range = Individual module temperature  
of 15 to 75 °C, as reported by the module, and  
environment temperature of 0 to 55 °C.  
Conversion type operating range  
Conversion types  
Auto  
Frequency range  
Controlled temperature range = Individual module  
temperature of 36 to 50 °C, as reported by the module,  
and environment temperature of 20 to 30 °C.  
All frequencies  
Double conversion  
Single high  
9 kHz to 3.6 GHz  
3.6 GHz to max frequency  
3.6 GHz to max frequency  
Specifications describe the warranted performance of  
calibrated instruments. Data represented in this document  
are specifications under the following conditions unless  
otherwise noted.  
Single low  
Additional information  
– Mixer level offset modifies the receiver gain prior to the  
first mixer of the receiver. A negative setting improves  
distortion (i.e., TOI) at the cost of noise performance  
(i.e., DANL). A positive setting improves noise  
performance at the cost of distortion.  
Calibrated instruments have been stored for a minimum  
of 2 hours within the full temperature range  
30 minute warm-up time  
Calibration cycle maintained  
– The PeakToAverage property is used with expected  
RF Power property to optimize level settings in the  
Downconverter. Set this to the ratio, in dB, of the peak  
power to the average power. The Downconverter uses  
this value to optimize mixer level, IF gain, and ADC  
clip level.  
When used with Keysight M9300A frequency reference  
and Keysight interconnect cables  
Characteristics describe product performance that is useful  
in the application of the product, but that is not covered by  
the product warranty. Characteristics are often referred to  
as Typical or Nominal values and are italicized.  
IF Level Offset (dB) provides additional adjustment of  
IF power level. Positive values reduce noise. Negative  
values reduce distortion.  
Typical describes characteristic performance, which  
80% of instruments will meet when operated within the  
controlled temperature range.  
Digitizer Level Offset (dB) provides additional adjustment  
of Downconverter IF power to the digitizer. Positive values  
increase power to the digitizer. Negative values decrease  
power to the digitizer.  
Nominal describes representative performance that is  
useful in the application of the product when operated  
within the controlled temperature range.  
– All graphs contain measured data from one unit and  
are representative of product performance within the  
controlled temperature range unless otherwise noted.  
95th percentile values indicate the breadth of the  
population (approx. 2 s of performance tolerances  
expected to be met in 95 percent of the cases with a  
95 percent confidence, for any ambient temperature in  
the range of 20 to 30 °C. In addition to the statistical  
observations of a sample of instruments, these values  
include the effects of the uncertainties of external  
calibration references. These values are not warranted.  
These values are updated occasionally if a significant  
change in the statistically observed behavior of  
production instruments is observed.  
Default conditions apply, unless otherwise noted.  
– The specifications contained in this document are  
subject to change.