23 | Keysight | M9393A PXIe Performance Vector Signal Analyzer – Data Sheet  
Technical Specifications and Characteristics  
Standard Configuration - Options F08, F14, F18, F27  
Format specific measurement data (cont’d)  
802.11g 64  
Parameters  
Nominal 1-channel  
EVM  
2.4 GHz, 20 MHz BW  
−50.5 dB  
802.11a 64  
Parameters  
Nominal 1-channel  
EVM  
5.8 GHz, 20 MHz BW  
−50 dB  
802.11n 64, 65  
Parameters  
64-QAM  
Nominal  
1-channel  
2-channel  
3-channel  
4-channel  
Preamble only  
EVM  
2.4 GHz, 40 MHz BW  
5.8 GHz, 40 MHz BW  
−48.4 dB  
−49.1 dB  
−47 dB  
−48 dB  
−47.9 dB  
−48.7 dB  
−50.5 dB  
Preamble, pilots, and data  
EVM  
2.4 GHz, 40 MHz BW  
5.8 GHz, 40 MHz BW  
−51.4 dB  
−52.2 dB  
−50.7 dB  
−51.8 dB  
−50.4 dB  
−51.2 dB  
802.11ac 64, 65  
Parameters  
256-QAM  
Nominal  
1-channel  
2-channel  
3-channel  
4-channel  
Preamble only  
EVM  
5.8 GHz, 80 MHz BW  
5.8 GHz, 160 MHz BW  
−48.5 dB  
−46 dB  
−46.9 dB  
−45.7 dB  
−45.5 dB  
−44.3 dB  
−46.4 dB  
−45.4 dB  
Preamble, pilots, and data  
EVM  
5.8 GHz, 80 MHz BW  
5.8 GHz, 160 MHz BW  
5.8 GHz, 80 MHz BW  
−51.5 dB  
−50.9 dB  
−49.4 dB  
−49.8 dB  
−46.9 dB  
−48.7 dB  
−47.1 dB  
−49.5 dB  
SEM  
See Figure 10  
802.11a/g 64  
Parameters  
SEM  
2.4 GHz, 20 MHz BW  
5.5 GHz, 20 MHz BW  
See Figure 11  
See Figure 12  
802.16e 64  
Parameters  
Nominal 1-channel  
OFDMA WiMAX™ EVM  
Fc = 2.5, 3.5, & 5.8 GHz  
−48 dB  
64. Expected input level 0 dBm, input signal (total power) 0 dBm, Mixer level offset 0 dB, conversion mode: Auto, PeakToAverage set per signal peak to  
average, demod symbol time adjustment –3.125%.  