06 | Keysight | M9393A PXIe Performance Vector Signal Analyzer – Data Sheet
Technical Specifications and Characteristics
Standard Configuration - Options F08, F14, F18, F27
Frequency
Frequency range and resolution
Option F08
Option F14
9 kHz to 8.4 GHz
9 kHz to 14 GHz
9 kHz to 18 GHz
9 kHz to 27 GHz
0.01 Hz
Option F18
Option F27
Tuning resolution
Analysis bandwidth 1
Maximum bandwidth
Option B04 (standard)
Option B10
40 MHz
100 MHz
Option B16
160 MHz
Option WB1 5
1 GHz IF output, nominal
IF frequency 2
Final IF
First IF (< 3.6 GHz)
5040 MHz
5100 MHz
5126 MHz
--
40 MHz IF path
240 MHz
300 MHz
326 MHz
Adjustable
100/160 MHz IF path
40 MHz alternate IF path 3
Bypass path (Option WB1)
6
Band
Harmonic mixing mode
LO multiple (N) 4
Frequency
Band 0
Band 1
Band 2
Band 3
Band 4
1
1
1
2
2
1
1
2
2
4
9 kHz to 3.6 GHz
3.6 to 8.4 GHz
8.4 to 13.6 GHz
13.6 to 17.1 GHz
17.1 to 27 GHz
1. Instantaneous bandwidth (1 dB bandwidth) available around a center frequency over which the input signal can be digitized for further
analysis or processing in the time, frequency or modulation domain.
2. Double conversion below 3.6 GHz, single conversion above 3.6 GHz.
3. Only used for some frequencies below 3.6 GHz for best performance as determined by the instrument software.
4. N is the LO multiplication factor.
5. Enables bypassing of IF filters to provide access to wideband IF output from downconverter for use with an external digitizer. Available > 3.6 GHz.
Full 1 GHz not available at band crossings.
6. IF frequency can be tuned from 200 to 800 MHz with a default value of 500 MHz.