29 | Keysight | M9393A PXIe Performance Vector Signal Analyzer – Data Sheet
Technical Specifications and Characteristics
Extended Frequency Configuration - Option FRX
Amplitude (cont’d)
Without M9169E
With M9169E
Amplitude switching speed
Trace detectors
—
Refer to M9169E data sheet
Same as standard configuration
Not available for this configuration
Preamplifier
Dynamic range
Displayed average noise level (DANL),
nominal 66
Without M9169E
With M9169E
3.6 to 13.6 GHz
> 13.6 to 34 GHz
> 34 to 45 GHz
> 45 to 50 GHz
–161 dBm/Hz
–158 dBm/Hz
–156 dBm/Hz
–153 dBm/Hz
–158 dBm/Hz
–153 dBm/Hz
–151 dBm/Hz
–147 dBm/Hz
Third order intermodulation distortion (TOI), Without M9169E
nominal
With M9169E
67
3.6 to 13.6 GHz
> 13.6 to 34 GHz
> 34 to 50 GHz
9.5 dBm / -53 dBc
5.5 dBm / -45 dBc
–2 dBm / -30 dBc
11 dBm / -56 dBc
8.5 dBm / -51 dBc
2.5 dBm / –39 dBc
Second harmonic distortion (SHI), nominal
3.6 to 13.6 GHz
Without M9169E
33 dBm / –33 dBc
25.5 dBm / –25.5 dBc
18 dBm / –18 dBc
With M9169E
Add loss from M9169E data sheet
13.6 to 34 GHz
34 to 50 GHz
74. Mixer level offset = 0, Conversion = auto, PeakToAverage = 0. Expected input level = -60 dBm
75. Tone separation = 100 kHz, Expected input level = -14 dBm, Mixer offset level = 0 dB, PeakToAverage = 6 dB, Conversion type Auto. Signal
level of -17 dBm used to calculate distortion in dBc. TOI = third order intercept. The TOI is given by the input tone level (in dBm) minus
(distortion/2) where distortion is the relative level of the distortion tones in dBc.