18 | Keysight | M9391A PXIe Vector Signal Analyzer - Data Sheet  
Description  
Specifications  
Supplemental Information  
Gain  
Instrument uncertainty53  
DUT gain range = -20 to +40 dB. See note54  
< 10 MHz  
10 MHz to 6 GHz  
± 0.21 dB  
53. “Instrument Uncertainty” is defined for gain measurements as uncertainty due to relative amplitude uncertainties encountered in the analyzer when making  
the measurements required for the gain computation. See Keysight App Note 57-2, literature number 5952-3706E for details on the use of this specification.  
Jitter (amplitude variations) will also affect the accuracy of results. The standard deviation of the measured result decreases by a factor of the square root  
of the Resolution Bandwidth used and by the square root of the number of averages. This application uses the 4 MHz Resolution Bandwidth as default since  
this is the widest bandwidth with uncompromised accuracy. Under difficult conditions (low Y factors), the instrument uncertainty for gain in high band can  
dominate the NF uncertainty as well as causing errors in the measurement of gain. These effects can be predicted with the uncertainty calculator.  
54. Uncertainty performance of the instrument is nominally the same in this frequency range as in the higher frequency range. However, performance is not  
warranted in this range. There is a paucity of available noise sources in this range, and the analyzer has poorer noise figure, leading to higher uncertainties  
as computed by the uncertainty calculator.  
Description  
Specifications  
Supplemental Information  
Noise figure uncertainty calculator55  
Instrument noise figure uncertainty  
Instrument gain uncertainty  
Instrument noise figure  
See the noise figure table earlier in this chapter  
See the gain table earlier in this chapter  
See graphs of “nominal instrument noise  
figure”; noise figure is DANL + 176.24 dB  
(nominal)56  
Instrument input match  
See graphs: nominal VSWR  
55. The Noise Figure Uncertainty Calculator requires the parameters shown in order to calculate the total uncertainty of a Noise Figure measurement.  
56. Nominally, the noise figure of the spectrum analyzer is given by NF = D — (K — L + N + B) where D is the DANL (displayed average noise level) specification,  
K is kTB (-173.98 dBm in a 1 Hz bandwidth at 290 K) L is 2.51 dB (the effect of log averaging used in DANL verifications) N is 0.24 dB (the ratio of the noise  
bandwidth of the RBW filter with which DANL is specified to an ideal noise bandwidth) B is ten times the base-10 logarithm of the RBW (in hertz) in which  
the DANL is specified. B is 0 dB for the 1 Hz RBW. The actual NF will vary from the nominal due to frequency response errors.  
Nominal instrument NF, 0.01 to 6 GHz, range = -28 dBm  
17  
16  
15  
14  
13  
12  
11  
10  
9
8
7
0
1
2
3
4
5
6
GHz  