04 | Keysight | M9391A PXIe Vector Signal Analyzer - Data Sheet  
Technical Specifications and Characteristics  
Definitions for specifications  
Conversion type operating range  
Temperatures referred to in this document are defined as follows:  
Full temperature range = Individual module temperature of 25  
to 75 °C, as reported by the module, and environment  
temperature of 0 to 55 °C.  
Controlled temperature range = Individual module tempera-  
ture of 40 to 51 °C, as reported by the module, and environ-  
ment temperature of 20 to 30 °C.  
Conversion types  
Auto  
Frequency range  
1 MHz to 3 or 6 GHz  
1 MHz to 3 or 6 GHz  
400 MHz to 3 or 6 GHz  
1.1 GHz to 3 or 6 GHz  
Image protect  
Single high  
Single low  
Additional information  
Specifications describe the warranted performance of calibrated  
instruments. Data represented in this document are specifications  
under the following conditions unless otherwise noted.  
Calibrated instruments have been stored for a minimum of 2  
hours within the full temperature range  
– 45 minute warm-up time  
Calibration cycle maintained  
When used with Keysight M9300A frequency reference and  
Keysight interconnect cables  
Mixer level offset modifies the receiver gain prior to the first  
mixer of the receiver. A negative setting improves distortion  
(i.e., TOI) at the cost of noise performance (i.e., DANL). A  
positive setting improves noise performance at the cost of  
distortion.  
Performance described in this document applies for module  
temperature within ± 3 degrees of comprehensive alignment,  
unless otherwise noted.  
When used with a Keysight M9018A PXIe chassis, compre-  
hensive alignment requires chassis FPGA version 1.05 or  
greater.  
Characteristics describe product performance that is useful in  
the application of the product, but that is not covered by the  
product warranty. Characteristics are often referred to as  
Typical or Nominal values and are italicized.  
Typical describes characteristic performance, which 80% of  
instruments will meet when operated  
within the controlled temperature range.  
Nominal describes representative performance that is useful  
in the application of the product when  
When configured for multi-channel, phase-coherent  
operation (shared synthesizer configuration), instrument level  
warranted specifications only apply to the M9391A which  
was previously calibrated with the M9301A synthesizer,  
showing a valid calibration indicator. For all other M9391A  
channels, specifications revert to typical performance. If  
using an external LO distribution unit, such as the V2802A LO  
distribution network, specifications for all M9391A channels  
revert to typical performance.  
All graphs contain measured data from one unit and is  
representative of product performance within the controlled  
temperature range unless otherwise noted.  
The specifications contained in this document are subject to  
change.  
operated within the controlled temperature range.  
Recommended best practices in use  
Use slot blockers and EMC filler panels in empty module slots  
to ensure proper operating temperatures. Keysight chassis  
and slot blockers optimize module temperature performance  
and reliability of test.  
Set chassis fan to high at environmental temperatures above  
45 °C  
Maintain temperature stability for best multi-channel phase  
coherence  
Set chassis fans to maximum  
Maintain stable ambient temperature  
Perform warm-up with session open and representative  
acquisition waveform running