23 | Keysight | M9391A PXIe Vector Signal Analyzer - Data Sheet
Technical Specifications and Characteristics
Environmental and physical specifications
Operating
Individual module temp 25 to 75 °C as reported by the module
and environment temp of 0 to 55 °C
Temperature
Non-operating (storage)
Environment temp of −40 to +70 °C
Type tested at 95%, +40 °C
(non-condensing)
Humidity 74
Operating random vibration
Survival random vibration
Functional shock
Bench handling
Type tested at 5 to 500 Hz, 0.21 g rms
Type tested at 5 to 500 Hz, 2.09 g rms
Type tested at half-sine, 30 g, 11 ms
Type tested per MIL-PRF-28800F
Shock/vibration 74
Up to 15,000 feet (4,572 meters) 75
Altitude
RF In
SMA female
Connectors
Complies with European EMC Directive 2004/108/EC
– IEC/EN 61326-2-1
– CISPR Pub 11 Group 1, class A
EMC
– AS/NZS CISPR 11
– ICES/NMB-001
This ISM device complies with Canadian ICES-001.
Cet appareil ISM est conforme a la norme NMB-001 du Canada.
45 minutes
Warm-up time
Size
M9300A
M9301A
M9350A
M9214A
1 PXIe slot
1 PXIe slot
1 PXIe slot
1 PXIe slot
Module
Length
Width
Height
Dimensions
M9300A
M9301A
M9350A
M9214A
210 mm
210 mm
210 mm
210 mm
22 mm
22 mm
22 mm
22 mm
130 mm
130 mm
130 mm
130 mm
M9300A
M9301A
M9350A
M9214A
0.55 kg (1.21 lbs)
0.54 kg (1.19 lbs)
0.56 kg (1.23 lbs)
0.36 kg (0.79 lbs)
Weight
M9300A
M9301A
M9350A
M9214A
≤ 18 W
≤ 25 W
≤ 30 W
≤ 35 W
Power drawn from chassis
74. Samples of this product have been type tested in accordance with the Keysight Environmental Test Manual and verified to be robust against the
environmental stresses of storage, transportation and end-use--those stresses include but are not limited to temperature, humidity, shock,
vibration, altitude and power-line conditions. Test methods are aligned with IEC 60068-2 and levels are similar to MIL-PRF-28800F Class 3.
75. At 15,000 feet, the maximum environmental temperature is de-rated to 52 °C.