23 | Keysight | M9391A PXIe Vector Signal Analyzer - Data Sheet  
Technical Specifications and Characteristics  
Environmental and physical specifications  
Operating  
Individual module temp 25 to 75 °C as reported by the module  
and environment temp of 0 to 55 °C  
Temperature  
Non-operating (storage)  
Environment temp of −40 to +70 °C  
Type tested at 95%, +40 °C  
(non-condensing)  
Humidity 74  
Operating random vibration  
Survival random vibration  
Functional shock  
Bench handling  
Type tested at 5 to 500 Hz, 0.21 g rms  
Type tested at 5 to 500 Hz, 2.09 g rms  
Type tested at half-sine, 30 g, 11 ms  
Type tested per MIL-PRF-28800F  
Shock/vibration 74  
Up to 15,000 feet (4,572 meters) 75  
Altitude  
RF In  
SMA female  
Connectors  
Complies with European EMC Directive 2004/108/EC  
– IEC/EN 61326-2-1  
– CISPR Pub 11 Group 1, class A  
EMC  
– AS/NZS CISPR 11  
– ICES/NMB-001  
This ISM device complies with Canadian ICES-001.  
Cet appareil ISM est conforme a la norme NMB-001 du Canada.  
45 minutes  
Warm-up time  
Size  
M9300A  
M9301A  
M9350A  
M9214A  
1 PXIe slot  
1 PXIe slot  
1 PXIe slot  
1 PXIe slot  
Module  
Length  
Width  
Height  
Dimensions  
M9300A  
M9301A  
M9350A  
M9214A  
210 mm  
210 mm  
210 mm  
210 mm  
22 mm  
22 mm  
22 mm  
22 mm  
130 mm  
130 mm  
130 mm  
130 mm  
M9300A  
M9301A  
M9350A  
M9214A  
0.55 kg (1.21 lbs)  
0.54 kg (1.19 lbs)  
0.56 kg (1.23 lbs)  
0.36 kg (0.79 lbs)  
Weight  
M9300A  
M9301A  
M9350A  
M9214A  
≤ 18 W  
≤ 25 W  
≤ 30 W  
≤ 35 W  
Power drawn from chassis  
74. Samples of this product have been type tested in accordance with the Keysight Environmental Test Manual and verified to be robust against the  
environmental stresses of storage, transportation and end-use--those stresses include but are not limited to temperature, humidity, shock,  
vibration, altitude and power-line conditions. Test methods are aligned with IEC 60068-2 and levels are similar to MIL-PRF-28800F Class 3.  
75. At 15,000 feet, the maximum environmental temperature is de-rated to 52 °C.  