13 | Keysight | M9391A PXIe Vector Signal Analyzer - Data Sheet
Technical Specifications and Characteristics
Dynamic range (continued)
Nominal Dynamic Range at 2 GHz
Pre-Amp Off, Image Protect
-50
-60
-70
-80
-90
-100
-110
-120
-130
-140
-150
-40
-35
-30
-25
-20
-15
-10
-5
Mixer Level (dBm) (Queried)
3rd Order Intermodulaꢀon
DANL (1Hz RBW)
DANL (30KHz RBW)
Using 15MHz IF Filters with 60MHz tone spacing for 3rd order intermodulaꢀon measurements
Figure 6. Dynamic range at 2 GHz, pre-amp OFF,
image protect conversion type.
Nominal Dynamic Range at 5.8 GHz
Pre-Amp Off, Image Protect
-50
-60
-70
-80
-90
-100
-110
-120
-130
-140
-150
-45
-40
-35
-30
-25
-20
-15
-10
Actual Mixer Level (dBm) (Queried)
3rd Order Intermodulaꢀon
DANL (1Hz RBW)
DANL (30KHz RBW)
Figure 7. Dynamic range at 5.8 GHz, pre-amp OFF,
image protect conversion type.