06 | Keysight | M9391A PXIe Vector Signal Analyzer - Data Sheet
Technical Specifications and Characteristics
Frequency
Frequency range and resolution
Option F03
1 MHz to 3 GHz
1 MHz to 6 GHz
0.001 Hz
Option F06
Tuning resolution
IF frequency
Nominal
326 MHz
240 MHz
300 MHz
15 MHz filter
40 MHz filter
160 MHz filter
Analysis bandwidth 1
Maximum bandwidth
Option B04
Option B10
Option B16
40 MHz
100 MHz
160 MHz
Frequency switching speed 2,3
Acquisition
bandwidth
Standard,
nominal
List mode switching speed 4
Sample rate
Option UNZ, nominal
Baseband frequency offset change 5
≤ 100 MHz
≤ 80 MHz
5 ms
5 ms
27 µs
> 100 MHz to
> 80 MHz to
102 µs
< 180 MHz
< 144 MHz
≥ 180 MHz
≥ 144 MHz
5 ms
5 ms
15 µs
Arbitrary frequency change
320 µs
Standard,
nominal
Non-list mode switching speed 6
Baseband frequency offset change 5
Arbitrary frequency change
Option UNZ, nominal
310 µs
5 ms
5 ms
2.3 ms
1.
Instantaneous bandwidth (1 dB bandwidth) available around a center frequency over which the input signal can be digitized for further
analysis or processing in the time, frequency or modulation domain.
2. When used with the M9018A PXIe chassis (2-link configuration: 1 x 8 [factory default]) and M9036A PXIe embedded controller.
3. Settled to within 1 kHz or 1 ppm, whichever is greater of final value. Does not include data acquisition or processing time. Amplitude
settled to within 0.1 dB. Channel filter set to none. Applies for all conversion types.
4. Time from trigger input to frequency and amplitude settled. Minimum IQ sample rate ≥ 6 MHz. Minimum spectrum acquisition ≥ 4.8 MHz.
Minimum power acquisition channel filter bandwidth ≥ 4.8 MHz. For lists with first point < 400 MHz or for frequency changes
from > 400 MHz to < 400 MHz, add 40 ms.
5. Baseband offset can be adjusted ± from carrier frequency within limits determined by RF analysis bandwidth and IF filter bandwidth.
Synthesizer frequency and amplitude are not changing. Baseband offset settled to within 1 kHz.
6. Mean time from IVI command to carrier frequency settled to within 1 kHz or 1 ppm, whichever is greater. Amplitude settled within 0.1 dB.
Simultaneous carrier frequency and amplitude switching. For frequency changes from > 400 MHz to < 400 MHz, add 40 ms.