05 | Keysight | M9381A PXIe Vector Signal Generator – Data Sheet  
Technical Specifications and Characteristics  
Additional information  
Definitions for specifications  
Specifications use the normal PLL mode setting, unless  
otherwise stated. Narrow loop bandwidth refers to  
Temperatures referred to in this document are defined as follows:  
specifications using the best wide offset PLL mode setting  
AGM938X_VAL_SYNTHESIZER_PLL_MODE_BEST_WIDE_  
OFFSET, available in the M938x Vector Signal Generator/CW  
Source Instrument Drivers versions 1.2.300 and later.  
Full temperature range = Individual module temperature  
of ≤ 75 °C as reported by the module, and environment  
temperature of 0 to 55 °C.  
Controlled temperature range = Individual module  
temperature of ≤ 55 °C as reported by the module, and  
environment temperature of 20 to 30 °C.  
Performance described in this document applies for mod-  
ule temperature within ± 5 degrees of IQ alignment, unless  
otherwise noted.  
Specifications describe the warranted performance of calibrated  
instruments. Data represented in this document are specifications  
under the following conditions unless otherwise noted.  
When configured for multi-channel, phase-coherent operation  
(shared synthesizer configuration), instrument level warranted  
specifications only apply to the M9381A which was previously  
calibrated with the M9301A synthesizer, showing a valid  
calibration indicator. For all other M9381A channels,  
specifications revert to typical performance. If using an ex-  
ternal LO distribution unit, such as the V2802A LO distribu-  
tion network, specifications for all M9381A channels revert to  
typical performance.  
Calibrated instruments have been stored for a minimum of  
2 hours within the full temperature range  
45 minute warm-up time  
Calibration cycle maintained  
When used with Keysight M9300A frequency reference and  
Keysight interconnect cables  
All graphs contain measured data from one unit and are  
representative of product performance within the controlled  
temperature range unless otherwise noted.  
Characteristics describe product performance that is useful in  
the application of the product. Characteristics are often referred  
to as Typical or Nominal values and are italicized.  
The specifications contained in this document are subject to  
changeꢀ  
Typical describes characteristic performance, which 80% of  
instruments will meet when operated within the controlled  
temperature range.  
Nominal describes representative performance that is useful  
in the application of the product when operated within the  
controlled temperature range.  
Recommended best practices in use  
Use slot blockers and EMC filler panels in empty module slots  
to ensure proper operating temperatures. Keysight chassis  
and slot blockers optimize module temperature performance  
and reliability of test.  
Set chassis fan to high at environmental temperatures  
above 45 °C.  
Maintain temperature stability for best multi-channel phase  
coherence  
Set chassis fans to maximum  
Maintain stable ambient temperature  
Perform warm-up with session open and representative  
waveform running