07 | Keysight | M9380A PXIe CW Source 1 MHz to 3 GHz or 6 GHz - Data Sheett
Technical Specifications and Characteristics (continued)
Amplitude
Output parameters
Settable range
Standard
Option 1EA
+10.7 to -130 dBm
+20 to -130 dBm
Resolution
ALC on 1
0.02 dB, nominal
0.3 dB, nominal
ALC off
Maximum output power
Frequency
Standard
+10 dBm
+10 dBm
Option 1EA
+19 dBm
+18 dBm
1 MHz to 5 GHz
> 5 to 6 GHz
Absolute level accuracy in CW mode [ALC on] 2
< Max power to -20 dBm
< -20 to -110 dBm
< -110 to -120 dBm
<-120 to -130 dBm
Frequency
1 MHz to 3 GHz
±0.4 dB
±0.5 dB
±0.7 dB
±0.8 dB, nominal
±0.15 dB, typical
±0.15 dB, typical
±0.25 dB, typical
> 3 to 6 GHz
±0.5 dB
±0.6 dB
±1.0 dB
±0.8 dB, nominal
±0.15 dB, typical
±0.25 dB, typical
±0.5 dB, typical
1. Settable to 0.01 dB.
2. Specifications apply at the controlled temperature range. For temperatures outside this range, absolute level accuracy degrades by ± 0.02 dB/°C.
-120 dBm Output Power Normalized Response
1
0.75
0.5
0.25
0
-0.25
-0.5
-0.75
-1
0
1
2
3
4
5
6
Frequency (GHz)
Figure 3. Output power normalized response at –120 dBm.