04 | Keysight | M9380A PXIe CW Source 1 MHz to 3 GHz or 6 GHz - Data Sheett
Technical Specifications and Characteristics
Definitions for specifications
Temperatures referred to in this document are defined as follows:
– Full temperature range = Individual module temperature of ≤ 75 °C, as reported by
the module, and environment temperature of 0 to 55 °C.
– Controlled temperature range = Individual module temperature of ≤ 55 °C, as
reported by the module, and environment temperature of 20 to 30 °C.
Specifications describe the warranted performance of calibrated instruments. Data
represented in this document are specifications unless otherwise noted under the
following conditions.
– Calibrated instruments have been stored for a minimum of 2 hours within the full
temperature range
– 45 minute warm-up time
– Calibration cycle maintained
– When used with Keysight M9300A frequency reference
– When used with Keysight interconnection cables
Characteristics describe product performance that is useful in the application of the
product. Characteristics are often referred to as Typical or Nominal values and are itali-
cized.
– Typical describes characteristic performance, which 80% of instruments will meet
when operated within the controlled temperature range.
– Nominal describes representative performance that is useful in the application of the
product when operated within the controlled temperature range.
Recommended best practices in use
– Use slot blockers and EMC filler panels in empty module slots to ensure proper
operating temperatures.
– Keysight chassis and slot blockers optimize module temperature performance and
reliability of test.
– At environment temperatures above 45 °C, chassis fan should be set to high.
Additional information
– All graphs contain measured data from one unit and is representative of product
performance at the controlled temperature range unless otherwise noted.
– The specifications contained in this document are subject to change.