04 | Keysight | M9380A PXIe CW Source 1 MHz to 3 GHz or 6 GHz - Data Sheett  
Technical Specifications and Characteristics  
Definitions for specifications  
Temperatures referred to in this document are defined as follows:  
Full temperature range = Individual module temperature of ≤ 75 °C, as reported by  
the module, and environment temperature of 0 to 55 °C.  
Controlled temperature range = Individual module temperature of ≤ 55 °C, as  
reported by the module, and environment temperature of 20 to 30 °C.  
Specifications describe the warranted performance of calibrated instruments. Data  
represented in this document are specifications unless otherwise noted under the  
following conditions.  
– Calibrated instruments have been stored for a minimum of 2 hours within the full  
temperature range  
– 45 minute warm-up time  
– Calibration cycle maintained  
– When used with Keysight M9300A frequency reference  
– When used with Keysight interconnection cables  
Characteristics describe product performance that is useful in the application of the  
product. Characteristics are often referred to as Typical or Nominal values and are itali-  
cized.  
– Typical describes characteristic performance, which 80% of instruments will meet  
when operated within the controlled temperature range.  
– Nominal describes representative performance that is useful in the application of the  
product when operated within the controlled temperature range.  
Recommended best practices in use  
Use slot blockers and EMC filler panels in empty module slots to ensure proper  
operating temperatures.  
– Keysight chassis and slot blockers optimize module temperature performance and  
reliability of test.  
At environment temperatures above 45 °C, chassis fan should be set to high.  