15 | Keysight | M9290A CXA-m PXIe X-Series Signal Analyzer - Data Sheet  
I/Q Analyzer  
Frequency  
Frequency span  
Standard instrument  
Option B25  
10 Hz to 10 MHz  
10 Hz to 25 MHz  
Resolution bandwidth (spectrum measurement)  
Range  
Overall  
100 MHz to 3 MHz  
Span = 1 MHz  
Span = 10 kHz  
Span = 100 Hz  
Window shapes  
50 Hz to 1 MHz  
1 Hz to 10 kHz  
100 MHz to 100 Hz  
Flat top, Uniform, Hanning, Gaussian, Blackman, Blackman-Harris, Kaiser Bessel (K-B 70 dB, K-B 90 dB and K-B 110 dB)  
Analysis bandwidth  
Standard instrument  
Option B25  
10 Hz to 10 MHz  
10 Hz to 25 MHz  
IF frequency response (standard 10 MHz IF path)  
IF frequency response (demodulation and FFT response relative to the center frequency, 20 to 30 °C)  
Center frequency (GHz)  
≤ 3.0  
Span (MHz)  
≤ 10  
Max. error  
± 0.40 dB  
RMS (nominal)  
0.03 dB  
3.0 < f ≤ 26.5  
≤ 10  
0.10 dB  
IF phase linearity (deviation from mean phase linearity, nominal)  
Center frequency (GHz)  
≤ 3.0  
3.0 < f ≤ 7.5  
Span (MHz)  
≤ 10  
≤ 10  
Peak-to-peak  
0.5 °  
0.5 °  
RMS  
0.2 °  
0.4 °  
Data acquisition (standard 10 MHz IF path)  
Time record length  
Sample rate  
4,000,000 IQ sample pairs  
90 MSa/s  
ADC resolution  
16 Bits  
Option B25 25 MHz analysis bandwidth  
IF frequency response (demodulation and FFT response relative to the center frequency, 20 to 30 °C)  
Center frequency (GHz)  
≤ 3.0  
3.0 < f ≤ 26.5  
Span (MHz)  
10 to ≤ 25  
10 to ≤ 25  
Max. error  
± 0.45 dB  
RMS (nominal)  
0.03 dB  
0.20 dB  
IF phase linearity (deviation from mean phase linearity, nominal)  
Center frequency (GHz)  
0.02 ≤ f < 3.0  
3.0 < f ≤ 7.5  
Span (MHz)  
10 to ≤ 25  
10 to ≤ 25  
Peak-to-peak  
1 °  
1 °  
RMS  
0.3 °  
0.5 °  
Data acquisition (B25 IF path)  
Time record length  
IQ analyzer  
4,000,000 IQ sample pairs  
90 MSa/s  
Sample rate  
ADC resolution  
16 Bits  
System Requirements  
Operating system  
Processor speed  
Windows 7 (32 & 64 bit)  
1.86 GHz minimum  
2.4 GHz recommended  
4 GB minimum  
Available memory  
8 GB recommended  
4 GB  
Available disk space  
Video  
Browser  
Support for DirectX 9 graphics with 128 MB graphics recommended (SuperVGA supported)  
Microsoft Internet Explorer 7.0 or greater  