12 | Keysight | M924XA InfiniiVision PXIe Modular Oscilloscopes - Data Sheet
Performance Characteristics
M924XA PXIe modular oscilloscopes
M924XA PXIe modular oscilloscopes overview
M9241A
M9242A
500 MHz
≤ 700 ps
2
M9243A
1 GHz
≤ 450 ps
2
Bandwidth (–3 dB) 1
200 MHz
≤ 1.75 ns
2
Calculated rise time (10 to 90%)
Input channels
Maximum sample rate
Maximum memory depth
Waveform update rate
5 GSa/s one channel, 2.5 GSa/s two channels
Standard 4 Mpts, standard segmented memory
≥ 1,000,000 waveforms/sec 2
Vertical system analog channels
M9241A
M9242A
M9243A
Hardware bandwidth limits
Input coupling
Approximately 20 MHz (selectable)
AC, DC
Input impedance
Input sensitivity range
Selectable: 1 MΩ ± 1% (15 pF), 50 Ω ± 3%
1 mV/div to 5 V/div (1 MΩ and 50 Ω)
1 mV/div to 5 V/div (1 MΩ)
1 mV/div to 1 V/div (50 Ω)
Vertical resolution
Maximum input voltage
8 bits (measurement resolution is 12 bits with averaging)
135 Vrms
With N2843A 10:1 probe: 300 Vrms
Frequency de-rating (assumes sine wave input): 400 Vpk until 40 kHz. Then de-rates at 20 db/dec
until 6 Vpk
DC vertical accuracy
± [DC vertical gain accuracy + DC vertical offset accuracy + 0.21% full scale] 3
± 2.0% full scale
± 0.1 div ± 2 mV ± 1% of offset setting
> 100:1 from DC to maximum specified bandwidth of each model (measured with same V/div and
coupling on channels)
DC vertical gain accuracy 1
DC vertical offset accuracy
Channel-to-channel skew
Offset range
± 2 V (1 mV/div to 200 mV/div)
± 50 V (> 200 mV/div to 5 V/div)
Time base range
Time base accuracy 1
2 ns/div to 50 s/div
1 ns/div to 50 s/div
500 ps/div to 50 s/div
Pre-trigger
± 1.6 ppm + aging factor (1st year: ± 0.5 ppm, 2nd year: ± 0.7 ppm, 5 years: ± 1.5 ppm, 10 years:
± 2.0 ppm)
Time base delay time range
Post-trigger Greater of 1 screen width or 250 μs
1 to 500 s
Channel-to-channel deskew range
∆ Time accuracy (using cursors)
Modes
± 100 ns
± (time base acc. x reading) ± (0.0016 x screen width) ± 100 ps
Main, zoom
1. Denotes warranted specifications. All others are typical.
2. Requires infinite persistence in order to visually display 1,000,000 wfm/sec.
3. Specifications are valid after a 30-minute warm-up period and ± 10 °C from firmware calibration temperature. 1 mV/div and 2 mV/div are a magnification of
4 mV/div setting. For vertical accuracy calculations, use full scale of 32 mV for 1 mV div and 2 mV/div sensitivity setting.