05 | Keysight | M9217A PXIe Single-Slot 2-Channel Isolated Digitizer - Data Sheet
Drivers
The M9217A digitizer is supplied with a comprehensive portfolio of module drivers, doc-
umentation, examples, and software tools to help you quickly develop test systems with
your software platform of choice. The module comes with IVICOM, IVI-C, and LabVIEW
G software drivers that work in the most popular test and measurement development
environments, including LabVIEW and LabWindows/CVI from National Instruments,
Microsoft C/C++, C#, and VB.NET.
Easy software integration
The module software support provides context-sensitive help, complete documentation
and code examples that allow a quick module set up and basic acquisition functionalities.
These code examples can be easily modified, so that the card can be quickly integrated
into a test system. Included are application code examples for LabVIEW, LabWindows/
CVI, Visual Studio C, C++, and C# and Visual Basic.
Calibration Intervals
The M9217A is factory-calibrated and shipped with a calibration certificate. Calibration is
recommended every year in order to verify product performance.
M9217A specifications
2-channel isolated digitizer with ADCs per channel and utilizing one PXI express slot.
Max sample rate
20 MSa/s
Sample resolution
16 bits
Input configuration
Isolated inputs ( each channel independently isolated)
+40 V
Isolation voltage (low to chassis)
Maximum input range (high to low) +256 Vpk
Input impedance
Input coupling
1 MW // 40 pF
DC or AC
< 10 Hz
AC cutoff frequency
Input ranges
±256 V, ±128 V, ±64 V, ±32 V, ±16 V, ±8 V,
±4V, ±2 V, ±1 V, ±500 mV, ±250 mV
Over voltage protection
Yes
Maximum overvoltage protection
Analog bandwidth (-3 dB)
±400 Vppk
20 MHz typical