17 | Keysight | M9195B PXIe Digital Stimulus/Response with PPMU: 250 MHz, 16-channel - Data Sheet
Technical specifications and characteristics
PPMU Characteristics
Characteristic
Value
Comments
* denotes warranted specification
Number of PPMU channels
PPMU modes
16
Force V measure V, Force V measure I, Force
I measure I, Force I measure V, Force nothing
measure V
Measurement averaging modes
Force voltage range
None, 64 averages, 50 Hz one PLC, 60 Hz one PLC
–2 V to +6.5 V (| Current | ≤ 4 mA)
–2 V to +6 V (| Current | ≤ 25 mA)
–2 V to +5.75 V (| Current | ≤ 40 mA)
(16-bit with 152 µV resolution)
Force voltage accuracy *
±10 mV
Maximum accuracy from ±5 °C of AutoCorrections.
Ground sense tied to ground.
Measure voltage accuracy *
±10 mV
With remote sense. Maximum accuracy from ±5 °C
of AutoCorrections with one PLC averaging at sense
location. Ground sense tied to ground
Force voltage settling time
< 20 µS (40 mA range)
< 20 µS (1 mA range)
< 25 µS (100 µA range)
< 100 µS (10 µA range)
< 525 µS (2 µA range)
1 V rising and falling step settled to 1% of final value
into 1 MΩ/ 1nF load. Typical
Force voltage stability
Current range
Stable at all ranges into 1 µF
Larger load capacitance possible, but response
limited by current slew rate.
–40 mA to + 40 mA
(16-bit with 2.44 µA resolution)
Maximum of 250 mA per module in combination
with other channels
–1 mA to + 1 mA
(16-bit with 61 nA resolution)
–100 µA to + 100 µA
(16-bit with 6.1 nA resolution)
–10 µA to + 10 µA
(16-bit with 610 pA resolution)
–2 µA to + 2 µA
(16-bit with 122 pA resolution)
Force current accuracy *
±1% of range
±1% of range
Maximum accuracy from ±5 °C of AutoCorrections.
Remote sense disconnected.
Measure current accuracy *
Maximum accuracy from ±5 °C of AutoCorrections
with one PLC averaging. Remote sense
disconnected.
Force current settling time
Channel leakage
Dependent on load
±10 nA
Typical
Remote sense
16 channels
1 GND
Ground sense should be tied to ground at
measurement location for maximum accuracy.
Remote sense leakage
±4 nA
Typical
Force current voltage clamp range
–2 V to +6.5 V
VCH > VCL
(16-bit with 152 µV resolution)
VCL (–2 V to +4 V)
VCH (0 V to +6.5 V)
Force current voltage clamp accuracy
±50 mV
Characteristic accuracy from ±5 °C of
AutoCorrections