13 | Keysight | M9195B PXIe Digital Stimulus/Response with PPMU: 250 MHz, 16-channel - Data Sheet  
Definitions for specifications and characteristics  
Specifications  
Warranted performance. Specifications include guardbands to account for the expected statistical performance distribution,  
measurement uncertainties, and changes in performance due to environmental conditions. All specifications and  
characteristics apply over the operating environment outlined in “Environmental and Regulatory” section of this data sheet. In  
addition, the following conditions must be met:  
Instrument has been turned on for 30 minutes with the PXI DSR software running.  
Instrument is within its calibration cycle.  
Instrument remains at a stable surrounding environment temperature (between 0°C to 45°C) for 1 hour prior to turn-on.  
Specifications in this document are identified by an asterisk (*).  
Characteristics  
Typical  
A performance parameter that the product is expected to meet before it leaves the factory, but that is not verified in the field. A  
characteristic includes the same guardbands as a specification.  
Expected performance of an average unit at a stable temperature between 20°C to 30°C for 30 minutes prior to turn-on and  
during operation; does not include guardbands. The instrument must be within its calibration cycle.  
Nominal  
A general, descriptive term or design parameter. It is not tested. Data represented in this document are nominal unless  
otherwise identified.  
Best practices  
To ensure proper cooling, use Keysight slot blockers (Y1212A), filler panels (Y1213A), and the air inlet module kit (Y1214A) in  
the chassis when there are empty slots. Keysight chassis and filler panels optimize module temperature performance.  
At environmental temperatures > 40 °C, chassis fans should be set to high.  
Additional information  
All data are measured from multiple units at room temperature and are representative of product performance within the  
operating temperature range unless otherwise noted.  
The specifications contained in this document are subject to change.  
Technical specifications and characteristics  
General Characteristics  
Module characteristics  
Bus interface & compatibility  
Number of data channels  
PXI Express peripheral module (x1, x4, x8 PCIe® specification v 2.1)  
16, per-channel parametric measurement unit (PPMU)  
Number of sites per module  
One 16 channel or four 4-channel sites (option dependent)  
Maximum data rate for data channels  
Maximum RTZ clock on data channels  
Number of high voltage channels  
Number of auxiliary open drain channels  
Module memory  
250 Mbps (option dependent)  
250 MHz (option dependent)  
4
4
2 GB (allocated between pattern, response capture and sequence control)  
Maximum number of synchronized modules 12 (requires opt MMS)  
Maximum number of synchronized channels 192 (requires opt MMS)  
Front panel connectors  
Data , open drain, HV, PPMU sense  
Reference clock input (REF CLK IN)  
Sync in/out (SYNC)  
ERCD30  
SMB connector  
SMB connector  
SMB connector  
Trigger in/marker out (GPIO1/2)  
Mechanical characteristics  
Dimensions  
3U/1-slot PXIe standard  
(H x W x D) in mm  
130.1 x 21.7 x 210 mm; includes connector and handle extensions  
Weight  
482 g  