19 | Keysight | M9195B PXIe Digital Stimulus/Response with PPMU: 250 MHz, 16-channel - Data Sheet
Technical specifications and characteristics
Environmental Characteristics
Operating and storage conditions
Temperature
Operating
Storage
0 to 45 °C
–40 to +70 °C
Humidity
Maximum 80% at 40°C (non-condensing)
Altitude
3,000 meters, de-rate max temperature by 5 °C above 2,000 meters
Calibration interval
Mechanical
1 year: return to Keysight service center or use Y1252A with N7800A/N7867A
Operating vibration
Survival sine vibration
Survival random vibration
Transportation shock
End use handling shock
Warm-up time
5-500 Hz: 0.21 g RMS, random
5-500 Hz: 0.5 g (0 to peak), swept sine
5-500 Hz: 2.09 g RMS, random
125 G, 8.6 m/s, trapezoidal pulse
1.6 m/s, <3 msec duration, half-sine pulse
30 minutes
Samples of this product have been type tested in accordance with the Keysight environmental test manual and verified to be robust against the
environmental stresses of storage, transportation and end-use; those stresses include but are not limited to temperature, humidity, shock, vibration,
altitude, and power line conditions. Test methods are aligned with IEC 60068-2 and levels are similar to MIL-PRF-28800F Class3.
Regulatory
Safety
IEC/ EN 61010-1
USA: ANSI/UL 61010
EMC
IEC 61326-1
Module calibration
Both adjustment and performance verification is required to calibrate M9195B modules. Self-maintainers can perform these
procedures by using software available from Keysight:
– N7800A Test management environment
– N7876A PXI digital IO calibration application
Module adjustment requires the Y1252A adjustment cable. This accessory consists of three cable assemblies necessary for voltage and
current adjustments of the M9195B. These cables are used with a high-end digital multimeter (DMM) to adjust on-board analog-digital
conversion. The N7876A is required to make adjustments using this cable.
Other instruments and hardware is required for calibration. The N7876A WebHelp file lists the equipment required for adjustment and
verification of the M9195B when using the N7800A and N7876A. For more details and access to the N7876A WebHelp file see:
http://cal.software.keysight.com/.