欢迎访问ic37.com |
会员登录 免费注册
发布采购

M9156CH40 参数 Datasheet PDF下载

M9156CH40图片预览
型号: M9156CH40
PDF下载: 下载PDF文件 查看货源
内容描述: [M9155/6/7C PXI Hybrid Switch Modules]
分类和应用:
文件页数/大小: 10 页 / 2839 K
品牌: KEYSIGHT [ Keysight Technologies ]
 浏览型号M9156CH40的Datasheet PDF文件第2页浏览型号M9156CH40的Datasheet PDF文件第3页浏览型号M9156CH40的Datasheet PDF文件第4页浏览型号M9156CH40的Datasheet PDF文件第5页浏览型号M9156CH40的Datasheet PDF文件第6页浏览型号M9156CH40的Datasheet PDF文件第8页浏览型号M9156CH40的Datasheet PDF文件第9页浏览型号M9156CH40的Datasheet PDF文件第10页  
07 | Keysight | M9155/6/7C PXI Hybrid Switch Modules DC to 26.5 GHz – Data Sheet  
Technical Specifications (continued)  
General operating data (continued)  
M9156C which contains two each of L7222C  
Hot switching  
Cold switching  
1 W CW, 50 W peak, 10 us max pulse width, not to exceed 1 W average  
150 W CW at 3 GHz, 25 °C, 120 W CW at 4.2 GHz, 25 °C  
M9157C which contains one L7106C  
Into internal termination  
Hot switching  
Into thru path  
Hot switching  
1 W CW, 50 W peak, 10 µs max pulse width, not to exceed 1 W average  
2 W CW, 100 W peak, 10 µs max pulse width, not to exceed 2 W average  
Environmental and physical specifications  
Temperature  
Operating  
0 to 55% C  
Non-operating  
–40 to 70% C  
Relative humidity  
Operating  
95% RF at 40% C, 24 hours cycling, repeated 5 times  
50% RH at –10% C to 25% C, 24 hour cycle  
Non-operating  
Vibration  
Operating random  
Vibration  
5 - 500 Hz, 0.3 g RMS  
5 - 500 Hz, 3.41 g RMS  
Survival random vibration  
Shock  
End use handling shock  
Transportation shock  
Half-sine wave form, 120 in/s, duration < 3 ms  
Trapezoidal, 50 g  
Altitude test  
Operating/non-operation  
15,000 ft (4600 m)  
ESD Immunity  
Air discharge  
15 kV per IEC61000-4-2  
8 kV per IEC61000-4-2  
Direct discharge1  
Safety  
EMC  
This product has input power below the requirements as specified in the Low Voltage Directive (2006/95/EC)  
EMC Standard: IEC 61326-1:2005/EN 61326-1:2006  
Emissions: CISPR 11:2003/EN55011:2007  
Immunity: IEC 61000-4-3:2002 / EN 61000-4-3:2002  
Electrostatic Discharge: IEC 61000-4-2:2001/EN 61000-4-2:1995+A1:1998+A2:2001  
EMC/EMI: CE, C-Tick  
CE compliance  
Warm-up time  
Dimensions  
EMC Compatibility Directive (EMC): 2004/108/EC  
Refer to PXI chassis warm up time  
3U PXI/CompactPCI standard  
Front panel complies with IEEE 1101.10 certification and compliance  
174.8 x 128.7 x 20 mm  
174.8 x 128.7 x 40.3 mm  
M9155C  
M9156C  
M9157C  
174.8 x 128.7 x 60.7 mm  
Weight  
M9155C  
M9156C  
M9157C  
0.395 kg  
0.255 kg  
0.340 kg  
Connectors  
PXI bus via 32 bit P1/J1 backplane connector. Signals via front panel mounted coaxial 3.5 mm (f) or SMA(f) connectors.  
Contact Material  
Beryllium copper, gold plated  
PXI-h, PXI-1, cPCI  
Connector compatibility  
1. To outer conductor