07 | Keysight | M9155/6/7C PXI Hybrid Switch Modules DC to 26.5 GHz – Data Sheet
Technical Specifications (continued)
General operating data (continued)
M9156C which contains two each of L7222C
Hot switching
Cold switching
1 W CW, 50 W peak, 10 us max pulse width, not to exceed 1 W average
150 W CW at 3 GHz, 25 °C, 120 W CW at 4.2 GHz, 25 °C
M9157C which contains one L7106C
Into internal termination
Hot switching
Into thru path
Hot switching
1 W CW, 50 W peak, 10 µs max pulse width, not to exceed 1 W average
2 W CW, 100 W peak, 10 µs max pulse width, not to exceed 2 W average
Environmental and physical specifications
Temperature
Operating
0 to 55% C
Non-operating
–40 to 70% C
Relative humidity
Operating
95% RF at 40% C, 24 hours cycling, repeated 5 times
50% RH at –10% C to 25% C, 24 hour cycle
Non-operating
Vibration
Operating random
Vibration
5 - 500 Hz, 0.3 g RMS
5 - 500 Hz, 3.41 g RMS
Survival random vibration
Shock
End use handling shock
Transportation shock
Half-sine wave form, 120 in/s, duration < 3 ms
Trapezoidal, 50 g
Altitude test
Operating/non-operation
15,000 ft (4600 m)
ESD Immunity
Air discharge
15 kV per IEC61000-4-2
8 kV per IEC61000-4-2
Direct discharge1
Safety
EMC
This product has input power below the requirements as specified in the Low Voltage Directive (2006/95/EC)
EMC Standard: IEC 61326-1:2005/EN 61326-1:2006
Emissions: CISPR 11:2003/EN55011:2007
Immunity: IEC 61000-4-3:2002 / EN 61000-4-3:2002
Electrostatic Discharge: IEC 61000-4-2:2001/EN 61000-4-2:1995+A1:1998+A2:2001
EMC/EMI: CE, C-Tick
CE compliance
Warm-up time
Dimensions
EMC Compatibility Directive (EMC): 2004/108/EC
Refer to PXI chassis warm up time
3U PXI/CompactPCI standard
Front panel complies with IEEE 1101.10 certification and compliance
174.8 x 128.7 x 20 mm
174.8 x 128.7 x 40.3 mm
M9155C
M9156C
M9157C
174.8 x 128.7 x 60.7 mm
Weight
M9155C
M9156C
M9157C
0.395 kg
0.255 kg
0.340 kg
Connectors
PXI bus via 32 bit P1/J1 backplane connector. Signals via front panel mounted coaxial 3.5 mm (f) or SMA(f) connectors.
Contact Material
Beryllium copper, gold plated
PXI-h, PXI-1, cPCI
Connector compatibility
1. To outer conductor