08 | Keysight | M9005A PXIe Chassis - Data Sheet
Recommended Configuration
Hardware
Configure the Keysight M9005A PXIe chassis as follows:
– Add the PC interface card
– The host adapter comes preconfigured with a 3 m PCIe
cable.
Model
Description
M9005A
M9005A-002
Includes
5-slot, PXIe chassis with integrated system module
PCIe Desktop adapter for M9005A + 3m cable
Standard PXI filler panels, getting started guide,
drivers, and Keysight I/O libraries
– Use slot blockers and EMC filler panels in empty module slots
to ensure proper module operating temperatures.
– Chassis fan may need to be set at high in order to provide
required module cooling at targeted environment cooling.
– Module cooling can be impacted by each module’s resistance
to air flow, therefore, refer to each module data sheet for
information regarding power and cooling to ensure proper
configurations.
– Place modules which have high cooling requirements in the
center slots of the M9005A. For high power modules, contact
your local Keysight office regarding compatibility of the
M9005A with your desired modular configuration.
Accessories
Model
Description
Y1212A
Y1213A
Y1274A
Slot blocker kit: 5 single-slot
PXI EMC filler panel kit: 5 single-slot filler panels
Rack mount kit
Services
Advantage Services: Calibration
Keysight Advantage Services is committed to your success throughout
your equipment’s lifetime.
Software information
Supported
Microsoft Windows 7 SP1 (32/64-bit)
operating systems Microsoft Windows 8.1 Update 1 (32/64-bit)
Microsoft Windows 10 (32/64-bit)
Standard
No IVI drivers required for this chassis
compliant drivers
Supported
application
development
environments
(ADE)
LabVIEW, LabWindows/CVI, MATLAB, VEE, Visual
Basic, VisualStudio.NET (C/C++, C#, VB.NET)
Keysight IO
Libraries
Supported versions: 17.2 (or later). Use of latest
version recommended. Includes: VISA Libraries,
Keysight Connection Expert, IO Monitor
Figure 2. Y1274A M9005A Rack Mount Kit.