04 | Keysight | E5061B Network Analyzer, 100 kHz to 1.5/3 GHz, 5 Hz to 500 M/1.5 G/3 GHz - Data Sheet
Table 2. Corrected system performance with Type-N 50 Ω connectors, 85032F calibration kit,
full 2-port calibration
Network analyzer: E5061B-3L3/3L4/3L5, calibration kit: 85032F (Type-N, 50 Ω), calibration: full 2-port
IF bandwidth = 10 Hz, No averaging applied to data, environmental temperature = 23 °C ± 5 °C with
< 1 °C deviation from calibration temperature, isolation calibration not omitted
Description
Specification (dB)
100 Hz to 100 kHz
100 kHz to 1 GHz
1 GHz to 3 GHz
Directivity
49
49
46
Source match
41
41
40
Load match
49
49
46
Reflection tracking
Transmission tracking
0.011
0.019
0.011
0.019
0.021
0.026
Figure 1. Transmission uncertainty (specification)
Figure 2. Reflection uncertainty (specification)