22 | Keysight | E5061B Network Analyzer, 100 kHz to 1.5/3 GHz, 5 Hz to 500 M/1.5 G/3 GHz - Data Sheet
Figure 20. Dynamic accuracy
Dynamic accuracy (Phase)
Dynamic accuracy (Magnitude)
10
100
10
1
1
0.1
0.01
0.1
-5
-15 -25 -35 -45 -55 -65 -75 -85 -95 -105 -115
-5 -15 -25 -35 -45 -55 -65 -75 -85 -95 -105 -115
Input power [dBm]
Input power [dBm]
Table 29. Test port input impedance
Description
Specification
Typical
Impedance
50 Ω nominal
1 MΩ / 30 pF
Return loss
> 15 dB at 50 Ω input
DC Bias (Option 3L3/3L4/3L5 only)
Table 30. DC bias
Description
DC voltage bias
Output Port
Range
Specification
Port 1, LF Out
0 to ± 40 V (100 mA max)
Resolution
1 mV ± (0 V to 10 V)
4 mV ± (10 V to 40 V)
Accuracy 1
± {0.1% + 4 mV} (at Open Port) 23 ± 5 °C
50 Ω nominal
Output Impedance
DC bias monitor
Voltage accuracy
at IFBW = AUTO (= < 100 Hz)
± {0.4% + 50 mV}
(at 23 °C ± 5 °C)
± {0.4% + 50 mV} x 4
(at 5 °C to 40 °C)
Automatic IF Bandwidth ≤ 100 Hz
Current accuracy
± {1% + 500 µA + (|Vdc[V]|/10 kΩ) }
(at 23 ± 5 °C)
± {1% + 500 µA + (|Vdc[V]|/10 kΩ) } x 2
(at 5 °C to 40 °C)
Automatic IF Bandwidth ≤ 100 Hz
1. DC Switching Transient Noise: ± 30 mV (SPD) when port or power switching occur.