03 | Keysight | E5061B Network Analyzer, 100 kHz to 1.5/3 GHz, 5 Hz to 500 M/1.5 G/3 GHz - Data Sheet
S-Parameter Measurement
Corrected system performance
The specifications in this section apply for measurements made with the Keysight
Technologies, Inc. E5061B network analyzer with the following conditions:
– No averaging applied to data
– Environmental temperature of 23 °C ± 5 °C, with less than 1 °C deviation from the
calibration temperature
– Response and isolation calibration not omitted
Table 1. System dynamic range 1,2
Description
Specification
SPD
System dynamic range
(Option 3L3/3L4/3L5)
100 kHz to 1 MHz, 3 kHz IF bandwidth
1 MHz to 3 GHz, 3 kHz IF bandwidth
5 to 100 Hz, 2Hz IF bandwidth
100 Hz to 9 kHz,10 Hz IF bandwidth
9 to 100 kHz, 10 Hz IF bandwidth
100 kHz to 1 MHz, 10 Hz IF bandwidth
1 MHz to 3 GHz, 10 Hz IF bandwidth
(Option 115, 135, 215, 235, 117, 137, 217, 237)
100 to 300 kHz, 3 kHz IF bandwidth
300 kHz to 1 MHz, 3 kHz IF bandwidth
1 MHz to 3 GHz, 3 kHz IF bandwidth
100 to 300 kHz, 10 Hz IF bandwidth
300 kHz to 1 MHz, 10 Hz IF bandwidth
1 MHz to 3 GHz, 10 Hz IF bandwidth
90 dB
95 dB
90 dB
100 dB
110 dB
115 dB
120 dB
130 dB
75 dB
90 dB
95 dB
100 dB
115 dB
120 dB
130 dB
1. The test port dynamic range is calculated as the difference between the test port rms noise floor and
the source maximum output power. The effective dynamic range must take measurement uncertainty
and interfering signals into account.
2. The specification might not be met at the frequencies 1.4 MHz, 4.0 MHz, 4.333 MHz, 6.167 MHz,
6.333 MHz, 25 MHz and 90 MHz.