Data Accuracy Enhancement
Measurement calibration
Measurement calibration significantly reduces measurement
uncertainty due to errors caused by system directivity, source
and load match, tracking and crosstalk. Full 2-port calibration
removes all the systematic errors for the related test ports to
obtain the most accurate measurements.
Calibration types available
Response
Simultaneous magnitude and phase correction of frequency
response errors for either reflection or transmission
measurements.
Response and isolation
Compensates for frequency response and crosstalk errors of
transmission measurements.
Enhanced response
One-port calibration
Compensates for frequency response and source match errors.
Compensates for directivity, frequency response and source
match errors.
Full 2-port calibration
Compensates for directivity, source match, reflection tracking,
(option 215, 235, 217, 237, 3L5) load match, transmission tracking and crosstalk. Crosstalk
calibration can be omitted.
Interpolated error correction
With any type of accuracy enhancement applied, interpolated
mode recalculates the error coefficients when the test frequencies
are changed. The number of points can be increased or decreased
and the start/stop frequencies can be changed.
Velocity factor
Enter the velocity factor to calculate the equivalent physical
length.
Reference port extension
Redefine the measurement plane from the plane where the
calibration was done.