32 | Keysight | E5061B Network Analyzer, 100 kHz to 1.5/3 GHz, 5 Hz to 500 M/1.5 G/3 GHz - Data Sheet
Source Control
Measured number of
points per sweep
User definable from 2 to 1601
Sweep type
Linear sweep, segment sweep, log sweep, power sweep and DC bias
sweep
Segment sweep
Define independent sweep segments. Set number of points, test port
power levels, IF bandwidth, delay time, sweep time independently for
each segment.
Sweep trigger
Power
Set to continuous, hold, or single, sweep with internal, external,
manual, or bus trigger.
Set source power from –45 dBm to 10 dBm. The power slope function
compensates source power error.
Trace Functions
Display data
Trace math
Title
Display current measurement data, memory data, or current
measurement and memory data simultaneously
Vector addition, subtraction, multiplication or division of measured
complex values and memory data
Add custom title to each channel window. Titles are printed on
hardcopies of displayed measurements
Autoscale
Electrical delay
Automatically selects scale resolution and reference value to vertically
center the trace
Offset measured phase or group delay by a defined amount of
electrical delay, in seconds
Phase offset
Statistics
Offset measured phase or group delay by a defined amount in degrees
Calculates and displays mean, standard deviation and peak-to-peak
deviation of the data trace