11 | Keysight | E5061B Network Analyzer, 100 kHz to 1.5/3 GHz, 5 Hz to 500 M/1.5 G/3 GHz - Data Sheet
Table 9. Corrected system performance with Type-N 75 Ω connectors, 85036B calibration kit,
enhanced response calibration
Network analyzer: E5061B-117/137/217/237, calibration kit: 85036B (Type-N, 75 Ω), calibration:
enhanced response calibration
IF bandwidth = 10 Hz, No averaging applied to data, environmental temperature
= 23 °C ± 5 °C with < 1 °C deviation from calibration temperature, isolation calibration not omitted
Description
Specification (dB)
100 kHz to 300 kHz 300 kHz to 1 MHz
1 MHz to 3 GHz
Directivity
49
48
44
Source match
48
41
35
Load match
48
48
44
Reflection tracking
Transmission tracking
0.004
0.022
0.010
0.028
0.019
0.052
Figure 15. Transmission uncertainty (specification)
Figure 16. Reflection uncertainty (specification)