26 | Keysight | E5061B Network Analyzer, 100 kHz to 1.5/3 GHz, 5 Hz to 500 M/1.5 G/3 GHz - Data Sheet  
Table 35. Analyzer environment and dimensions  
Description  
General characteristics  
Operating environment  
Temperature  
+5 °C to +40 °C  
Error-corrected temperature range  
23 °C ± 5 °C with < 1 °C deviation from calibration  
temperature  
Humidity  
20% to 80% at wet bulb temperature < +29 °C  
(non-condensing)  
Altitude  
0 to 2,000 m (0 to 6,561 feet)  
Vibration  
0.21 G maximum, 5 Hz to 500 Hz  
Non-operating storage environment  
Temperature  
–10 °C to +60 °C  
Humidity  
20% to 90% at wet bulb temperature < 40 °C  
(non-condensing)  
Altitude  
0 to 4,572 m (0 to 15,000 feet)  
0.5 G maximum, 5 Hz to 500 Hz  
Vibration  
Dimensions  
Height = 235 mm, Width = 432 mm, Depth = 296 mm.  
See Figures 10 to 12.  
Weight  
14.0 kg  
Magnetic susceptibility  
Degradation of some product specifications can occur in  
the presence of ambient power frequency magnetic fields of  
30 A/m or greater.  
The product self-recovers and operates as specified when  
removed or shielded from the ambient magnetic field.  
When the analyzer tuned frequency is identical to the  
immunity test signal frequency, there may be signals of up to  
−80 dB of full-scale response displayed on the screen.  
Magnetic emission  
Emission of magnetic field may occur at the left side of the  
where two cooling fans are installed. Its magnitude can be  
as much as 160 A/m and 25 A/m at 0 cm and 1 cm apart  
from the center of the fan, respectively. It is recommended  
to have enough clearance between the cooling fans and  
magnetically sensitive device or instruments.  