31 | Keysight | E5061B Network Analyzer, 100 kHz to 1.5/3 GHz, 5 Hz to 500 M/1.5 G/3 GHz - Data Sheet
Measurement Capabilities
Number of measurement
channels
Up to 4 independent measurement channels. A measurement
channel is coupled to stimulus response settings including
frequency, IF bandwidth, power level, and number of points.
Number of display windows
Each measurement channel has a display window. Up to
4 display windows (channels) can be displayed.
Number of traces
4 data traces and 4 memory traces per channel
Measurement choices
Option 115,135,117 and 137 - S11, S21, Absolute
Option 215, 235 and 237 - S11, S21, S12, S22, Absolute
Option 3L5 - S11, S21, S12, S22, T/R, T, R, Absolute
Measurement parameter
conversion
Available to convert S-parameters into reflection impedance,
transmission impedance, reflection admittance, transmission
admittance, and 1/S.
Data formats
Data markers
Log magnitude, linear magnitude, phase, expanded phase,
positive phase, group delay, SWR, real, imaginary, Smith chart,
polar.
10 independent markers per trace. Reference marker available
for delta marker operation. Smith chart format includes
5 marker formats: linear magnitude/phase, log magnitude/
phase, real/imaginary, R + jX, and G + jB. Polar chart format
includes 3 marker formats: linear magnitude/phase, log
magnitude/phase, and real/imaginary.
Marker functions
Marker search
Max value, min value, multi-peak, multi-target, peak, peak left,
peak right, target, target left, target right, and width parameters
with user-defined bandwidth values.
Marker-to functions
Set start, stop, center to active marker stimulus value; set
reference to active marker response value; set electrical delay
to group delay at active marker.
Search range
Tracking
User definable.
Performs marker search continuously or on demand.
Fault location functions (Option 010)
Transformation to distance and
time domain
Selectable transformation type from bandpass, lowpass
impulse, lowpass step.
Selectable window from maximum, normal and minimum.
Impedance measurement analysis (Option 005)
Impedance Measurement Selectable Impedance Parameter and Equivalent Circuit analysis
Capability
Wireless power transfer analysis (Option 006) 1
Wireless power transfer circuit
evaluation
Selectable parameters related to wireless power transfer
analysis. 2D/3D simulation capability.
1. Option 006 can be installed with option 215/235/3L5 only