13 | Keysight | E5061B Network Analyzer, 100 kHz to 1.5/3 GHz, 5 Hz to 500 M/1.5 G/3 GHz - Data Sheet
Test Port Output (Source)
Table 11. Test port output frequency
Description
Specification
5 Hz to 500 MHz
5 Hz to 1.5 GHz
5 Hz to 3 GHz
100 kHz to 3 GHz
100 kHz to 1.5 GHz
1 mHz
Typical
Range (Option 3L3)
Range (Option 3L4)
Range (Option 3L5)
Range (Option 115, 135, 215, 235, 137, 237)
Range (Option 135, 235, 137, 237)
Resolution
Source stability
± 7 ppm (5 to 40 °C)
CW accuracy
High stability option (Option 1E5)
CW accuracy
± 7 ppm ± 1 mHz
± 1 ppm ± 1 mHz
Stability
± 0.05 ppm (5 to 40 °C)
± 0.5 ppm per year
Table 12. Test port output power
Description
Specification
Typical
Level accuracy
(Option 3L3/3L4/3L5)
± 0.8 dB (at 0 dBm, 50 MHz absolute)
± 1.0 dB (at 5 Hz to 1.5 GHz, 0 dBm,
relative to 50 MHz)
± 1.5 dB (at 1.5 GHz to 3 GHz, 0 dBm,
relative to 50 MHz)
Level accuracy
(Option 115, 135, 215,
235)
± 0.8 dB (at 0 dBm, 50 MHz absolute)
± 1.5 dB (at 100 kHz to 300 kHz, 0 dBm,
relative to 50 MHz)
± 1.0 dB (at 300 kHz to 3 GHz, 0 dBm,
relative to 50 MHz)
Level accuracy
(Option 117, 137, 217,
237)
± 0.8 dB (at 0 dBm, 50 MHz absolute)
± 1.5 dB (at 100 kHz to 300 kHz, 0 dBm,
relative to 50 MHz)
± 1.0 dB (at 2 GHz to
3 GHz, 0 dBm, relative
to 50 MHz)
± 1.0 dB (at 300 kHz to 2 GHz, 0 dBm,
relative to 50 MHz)
Level lineality
± 0.75 dB (at –10 to 10 dBm, 0 dBm reference)
(Option 3L3/3L4/3L5)
Level lineality
(Option 115, 135, 215,
± 0.75 dB (at –10 to 5 dBm, 100 kHz to 300 kHz,
0 dBm reference)
235, 117, 137, 217, 237) ± 0.75 dB (at –10 to 10 dBm, 300 kHz to 3 GHz,
0 dBm reference)
Range
–45 dBm to 10 dBm
(Option 3L3/3L4/3L5)
Range
–45 dBm to 5 dBm (at 100 kHz to 300 kHz)
–45 dBm to 10 dBm (at 300 kHz to 3 GHz)
(Option 115, 135, 215,
235, 117, 137, 217, 237)
Sweep range
–45 dBm to 10 dBm
(Option 3L3/3L4/3L5)
Sweep range
–45 dBm to 5 dBm (at 100 kHz to 300 kHz)
–45 dBm to 10 dBm (at 300 kHz to 3 GHz)
(Option 115, 135, 215,
235, 117, 137, 217, 237)
Level resolution
0.05 dB
Table 13. Test port output signal purity
Description
Specification
Typical
Harmonics (2nd or 3rd)
< -20 dBc (at 100 kHz to 300 kHz, 5 dBm)
< -25 dBc (at 300 KHz to 3 GHz, 5 dBm)
< –25 dBc (at 5 dBm)
Non-harmonic spurious