21 | Keysight | E5061B Network Analyzer, 100 kHz to 1.5/3 GHz, 5 Hz to 500 M/1.5 G/3 GHz - Data Sheet
Table 26. Test port input (trace noise)
Description
Specification
Typical
Trace noise
(at IF automatic bandwidth, < 10 kHz)
5 mdB rms
5 mdB rms
(at 3 kHz bandwidth, 10 kHz to
30 MHz at –5 dBm, 0 dB attenuation, 50 Ω)
Trace noise phase
(at IF automatic bandwidth, < 10 kHz)
0.03 ° rms
0.03 ° rms
(at 3 kHz bandwidth, 10 kHz to
30 MHz at –5 dBm,
0 dB attenuation, 50 Ω)
Table 27. Test port input (stability)
Description
Specification
SPD
Stability magnitude
Stability phase
< ± 0.02 dB/°C
< ± 0.2 °/°C
Table 28. Test port input (Dynamic accuracy) 1
Description
Specification
Typical
Dynamic accuracy magnitude
(0 dB attenuation, 50 Ω)
± 0.303 dB at –5 dBm
± 0.09 dB at –15 dBm
± 0.056 dB at –25 dBm
± 0.073 dB at –35 dBm
± 0.087 dB at –45 dBm
± 0.103 dB at –55 dBm
± 0.121 dB at –65 dBm
± 0.15 dB at –75 dBm
± 0.211 dB at –85 dBm
± 0.371 dB at –95 dBm
± 0.841 dB at –105 dBm
± 2.141 dB at –115 dBm
Dynamic accuracy phase
(0 dB attenuation, 50 Ω)
± 2.04 ° at –5 dBm
± 0.6 ° at –15 dBm
± 0.37 ° at –25 dBm
± 0.48 ° at –35 dBm
± 0.58 ° at –45 dBm
± 0.68 ° at –55 dBm
± 0.81 ° at –65 dBm
± 1.00 ° at –75 dBm
± 1.41 ° at –85 dBm
± 2.5 ° at –95 dBm
± 5.83 ° at –105 dBm
± 16.23 ° at –115 dBm
± 5 °
(+15 dBm, 20 dB
attenuation)
1. Accuracy of the test port input power reading is relative to –25 dBm reference input power level.