02 | Keysight | E4991B Impedance Analyzer - Data Sheet  
Definitions  
Specification (spec.)  
Warranted performance. All specifications apply at 23 °C ± 5 °C unless otherwise stated,  
and 30 minutes after the instrument has been turned on. Speciications include guard  
bands to account for the expected statistical performance distribution, measurement  
uncertainties, and changes in performance due to environmental conditions.  
Typical (typ.)  
Expected performance of an average unit which does not include guardbands. It is not  
covered by the product warranty.  
Nominal (nom.)  
A general, descriptive term that does not imply a level of performance. It is not covered  
by the product warranty.  
Measurement Parameters and Range  
Measurement parameters  
Impedance parameters:  
|Z|, |Y|, Ls, Lp, Cs, Cp, Rs(R), Rp, X, G, B, D, Q, qz, Gy, |G|, Gx, y, qr, Vac, Iac, Vdc, Idc  
(Option E4991B-001 only)  
Material parameters (Option E4991B-002):  
(see “Option E4991B-002 material measurement (typical)” on page 19)  
Permittivity parameters: |r|, r’, r”, tan  
Permeability parameters: |r|, r’, r”, tan  
Measurement range  
Measurement range (|Z|):  
120 mΩ to 52 kΩ.  
(Frequency = 1 MHz,  
Point averaging factor ≥ 8,  
Oscillator level = –3 dBm; or = –13 dBm,  
Measurement accuracy ≤ ± 10%,  
Calibration is performed within 23 °C ± 5 °C,  
Measurement is performed within ± 5 °C of calibration temperature)  
Source Characteristics  
Frequency  
Range: 1 MHz to 3 GHz (Option 300)  
1 MHz to 1 GHz (Option 100)  
1 MHz to 500 MHz (Option 050)  
Resolution: 1 mHz  
Accuracy:  
without Option E4991B-1E5:  
± 10 ppm (23 °C ± 5 °C)  
± 20 ppm (5 °C to 40 °C)  
with Option E4991B-1E5:  
± 1 ppm (5 °C to 40 °C)  
 复制成功!