02 | Keysight | E4991B Impedance Analyzer - Data Sheet
Definitions
Specification (spec.)
Warranted performance. All specifications apply at 23 °C ± 5 °C unless otherwise stated,
and 30 minutes after the instrument has been turned on. Speciications include guard
bands to account for the expected statistical performance distribution, measurement
uncertainties, and changes in performance due to environmental conditions.
Typical (typ.)
Expected performance of an average unit which does not include guardbands. It is not
covered by the product warranty.
Nominal (nom.)
A general, descriptive term that does not imply a level of performance. It is not covered
by the product warranty.
Measurement Parameters and Range
Measurement parameters
Impedance parameters:
|Z|, |Y|, Ls, Lp, Cs, Cp, Rs(R), Rp, X, G, B, D, Q, qz, Gy, |G|, Gx, y, qr, Vac, Iac, Vdc, Idc
(Option E4991B-001 only)
Material parameters (Option E4991B-002):
(see “Option E4991B-002 material measurement (typical)” on page 19)
Permittivity parameters: |r|, r’, r”, tan
Permeability parameters: |r|, r’, r”, tan
Measurement range
Measurement range (|Z|):
120 mΩ to 52 kΩ.
(Frequency = 1 MHz,
Point averaging factor ≥ 8,
Oscillator level = –3 dBm; or = –13 dBm,
Measurement accuracy ≤ ± 10%,
Calibration is performed within 23 °C ± 5 °C,
Measurement is performed within ± 5 °C of calibration temperature)
Source Characteristics
Frequency
Range: 1 MHz to 3 GHz (Option 300)
1 MHz to 1 GHz (Option 100)
1 MHz to 500 MHz (Option 050)
Resolution: 1 mHz
Accuracy:
without Option E4991B-1E5:
± 10 ppm (23 °C ± 5 °C)
± 20 ppm (5 °C to 40 °C)
with Option E4991B-1E5:
± 1 ppm (5 °C to 40 °C)