19 | Keysight | E4991B Impedance Analyzer - Data Sheet  
Option E4991B-002 Material Measurement (Typical)  
Measurement parameter  
Permittivity parameters: |[r|, [r’, [r”, tand  
Permeability parameters: |µr|, µr’, µr”, tand  
Frequency range  
Using with Keysight Technologies, Inc. 16453A:  
1 MHz to 1 GHz (typical)  
Using with Keysight 16454A: 1 MHz to 1 GHz (typical)  
Measurement accuracy  
Conditions for defining accuracy:  
Calibration:  
Open, short, and load calibration at the fixture (7-mm connector)  
Calibration temperature:  
Calibration is performed at an environmental temperature within the range of  
23 °C ± 5 °C.  
Measurement accuracy doubles when calibration temperature is 5 °C to 18 °C or  
28 °C to 40 °C.  
Temperature:  
Temperature deviation: within ± 5 °C from the calibration temperature  
Environment temperature: Measurement accuracy applies when the calibration is  
performed at 23 °C ± 5 °C. When the calibration is below 18 °C or above 28 °C,  
measurement error doubles.  
Measurement frequency points:  
Same as calibration points1  
Point averaging factor: ≥ 8  
Electrode pressure setting of 16453A: maximum  
Typical accuracy of permittivity parameters:  
Δ[' r m  
[raccuracy  
=
:
'
[
r m  
0.1  
t
'
['r m  
100  
13  
± 5 + 10 +  
+ 0.25  
r m  
+
[%]  
t
2
f
[
1–  
f ['  
r m  
(at tand < 0.1)  
Loss tangent accuracy of [r (= tand):  
± (Ea + Eb) (at tand < 0.1)  
where,  
Ea  
=
at Frequency ≤ 1 GHz:  
0.001  
t
0.1  
13  
0.002 +  
+ 0.004f +  
f
['  
2
r m  
1–  
f ['  
r m  
1. In fixed frequency calibration mode, if a measurement frequency point is not included in the  
calibration points, the accuracy at the measurement point is degraded to its doubled value (typical).  