06 | Keysight | E4990A Impedance Analyzer 20 Hz to 10/20/30/50/120 MHz - Data Sheet  
DC Bias Function  
DC voltage bias  
Range  
0 to ± 40 V (see Figure 1)  
1 mV  
Resolution  
Accuracy  
± [0.1% + (5 + 30 × |Imon|) mV]  
± [0.2% + (10 + 30 × |Imon|) mV] (beyond 23 ± 5 °C)  
DC current bias  
Range  
0 to ± 100 mA (see Figure 1)  
40 µA  
Resolution  
Accuracy  
± [2% + (0.2 + |Vmon|/20) mA]  
± [4% + (0.4 + |Vmon|/20) mA] (beyond 23 ± 5 °C)  
DC voltage bias at constant voltage mode  
Range  
0 to ± 40 V (see Figure 1)  
1 mV  
Resolution  
Accuracy  
± [0.5% + (5 + Zd × |Imon|) mV] (typical)  
± [1.0% + (10 + Zd × |Imon|) mV] (beyond 23 ± 5 °C, typical)  
DC current bias at constant current mode  
Range  
0 to ± 100 mA (see Figure 1)  
40 µA  
Resolution  
Accuracy  
± [1% + (0.5 + |Vmon|/10000) mA] (typical)  
± [2% + (1.0 + |Vmon|/5000) mA] (beyond 23 ± 5 °C, typical)  
DC bias monitor  
DC voltage range  
DC voltage accuracy  
(Same as the dc voltage bias setting range)  
± [0.2% + (5 + Zd × |Imon|) mV]  
± [0.4% + (10 + Zd × |Imon|) mV] (beyond 23 ± 5 °C)  
DC current range  
(Same as the dc voltage bias setting range)  
DC current monitor accuracy  
± [1% + (0.5 + |Vmon|/10000) mA]  
± [2% + (1.0 + |Vmon|/5000) mA] (beyond 23 ± 5 °C)  
Output impedance  
25 Ω (nominal)  
NOTE  
Vmon : DC voltage bias monitor reading value [mV]  
Imon  
Zd  
Zd  
Zd  
Zd  
Zd  
: DC current bias monitor reading value [mA]  
= 0.3 (at four-terminal pair port of the E4990A, adapter setup: NONE)  
= 2.0 (at test port of the 42941A, adapter setup: 42941A Impedance Probe)  
= 0.5 (at 7-mm port of the 42942A, adapter setup: 42942A Terminal Adapter)  
= 1.0 (at measurement port of the 16048G, adapter setup: four-terminal pair 1 m)  
= 1.5 (at measurement port of the 16048H, adapter setup: four-terminal pair 2 m)  
Figure 1. DC bias range  